well, and you can also set and change the wells to contain different sample types, patient sample, controls, calibrators

or empty wells. You cannot alter wells with spots already punched into them.
To select a single well, press it. To select multiple wells, drag your fingers across the plate to select an area containing

the wells you want. The wells are selected based on the filling order (row or column). To select a single column, press

the corresponding number and, similarly, if you want to select a row, press the corresponding letter. To select the whole

plate with a single touch, press the upper left corner of the plate.
The calibrator button opens a sub-view where you can view and select all calibrator levels. There is a similar

functionality for controls too. The 

Patients

 sample type represents unknown samples and the 

Empty

 type represents

wells that will not be used at all for punching. The 

Punched

 type represents samples that have been punched outside

the instrument or for instance a liquid control that you want the system to show in the worklist. If a well is marked as

Punched

, it will be displayed as a black disk on a white background until it has been assigned an ID.

Note: 

The well will appear as empty in the worklist if no ID has been assigned to it.

Note: Punched

 is not to be used when attached to Specimen Gate Laboratory. Using it will cause wells to appear as

empty in Specimen Gate Laboratory.
Greyed out wells in this view belong to the explanatory classes:

Class

Explanation

Repunch sample

For samples that will be punched to the same well again.

Bad well

Wells that contain some kind of error.

Start location

The first patient punching location, if not the first patient

location. This may be set for a plate through the loading

template.

If you select a well that has already been punched into, the sample info is displayed on the right side of the plate map.

Note: 

Editing a punched well requires you to be in 

Check

 mode under the 

Punch

 view.

If you want to export the plate data manually, press the 

Export as

 button and in the dialog that appears, select the

export file type and location where you want to save it.

Note: 

Only one folder path is stored for manual exporting and importing. If you have imported a file previously and

you now want to export data, the default folder path is that of the file previously imported. If needed, remember to

change the folder path to point to the correct location.
If a plate needs to be reloaded to the instrument, you can recover its data either from the recovery file or from a

panthera.xml type export file, if available. To enable plate data import, the plate must have the same plate barcode, the

same analyte, and it cannot contain any new disk, i.e punched after it was unloaded. There are many folders where the

plate data can be located, so if you cannot immediately see the data file, please check the following folders: panthera

exports and its subdirectories, panthera recovery or any other possible export locations you might have set. Clicking

the path-input box will open a directory browser that provides quick links to the Panthera export and Panthera recovery

locations.
To pack all the wells on the plate, press 

Pack

. This makes it so the controls appear after the last punched patient well.

For plates that do not have end controls, packing removes all the empty wells from the plate map. The 

Pack

 function

works as described above in stand-alone mode. With Specimen Gate Laboratory, 

Pack

 works only if the assay has been

programmed to have End Controls in Specimen Gate.
If for any reason you need to assign a note to a particular well, select that well and press inside the 

Note

 field on the

right of the plate map. The software will show you the 

Note

 screen, in which you can either enter text (free form) into

the 

Current note

 field or select one of the previously entered notes. The software remembers the last 12 notes that

have been entered. If needed, PerkinElmer personnel can lock notes in the list for you, so that they are never removed,

even if they have not been used for some time. The total number of notes allowed by the software is 12 regardless of

whether they are locked or first in first out.

Note: 

Notes can only be added to wells already containing a sample unless a note is selected prior to punching. If a

note was selected prior to punching, the note will be attached to each well that is punched.

Note: 

When working with Specimen Gate, for notes to be flagged as either informational or quality flags, the

notes entered into the Panthera software must match the notes programmed in Specimen Gate. When connected

to Specimen Gate, notes cannot be removed, but they can be updated.

Panthera-Puncher

 9 User Manual

Routine operation

11

Содержание Panthera-Puncher 9

Страница 1: ...2081 0010 USER MANUAL Panthera Puncher 9 Software version 1 8 Dried Blood Spot Punching Device...

Страница 2: ......

Страница 3: ...and setting up the plates 9 Unloading the plates from the instrument 10 Plate edit view 10 Operations in detail 12 Punching and sample queries 12 Sample identification 13 Detection of punch areas 13...

Страница 4: ...25 Loading templates 25 Analytes 25 Punching 26 Punches per well 26 Card barcode rules 27 Plate 27 Exporting 27 Plate maps 28 Barcode rule notation 28 Storing of the sample card image 30 Log files 31...

Страница 5: ...t from the Keyboards and Languages tab of the Windows Region and Language dialog The default user and administrator user names and corresponding passwords are provided in a separate document If the ne...

Страница 6: ...of moving parts in the instrument This way up Fragile handle with care Keep dry Certification mark from testing laboratory In vitro diagnostic medical device Catalogue number Battery recycle WEEE foll...

Страница 7: ...heck the number of disks in the well in error or repunch situations Incorrect number of disks may lead to a false result Warning Installing the instrument in a location where sunlight reaches the samp...

Страница 8: ...eads Caution When cleaning the punch heads ensure that parts from different punch heads do not become interchanged This may cause the punch head to become unusable the disk to not have space to fall o...

Страница 9: ...need to be added to certain user groups in order to have access rights to the Panthera software User groups in question are Puncher Users or Puncher Administrators depending on what functions the use...

Страница 10: ...nched from each blood spot Intellipunch detects the blood areas on the sample card based on the card surface color and automatically optimizes the punch locations for the current blood spot This provi...

Страница 11: ...t able to recover the error message will say Unrecoverable error When the instrument informs the user of e g a movement error it is a hint that an error has occurred but that the instrument was able t...

Страница 12: ...all controls 6 green dots to the plate prior to patient samples By default it is set to punch only the start controls so that the plate can be packed in case it is a partial plate It is not possible...

Страница 13: ...cause the plates to become unusable Loading and setting up the plates Press the Eject button in the Plates view and open the hatch Once the hatch is open the plate track comes out The plate track cont...

Страница 14: ...head configuration When connected to Specimen Gate Laboratory all samples with the unavailable disk size will still be displayed on the punch screen albeit they will be disabled and you will not be a...

Страница 15: ...now want to export data the default folder path is that of the file previously imported If needed remember to change the folder path to point to the correct location If a plate needs to be reloaded to...

Страница 16: ...er the first punch command realign the card and the instrument will locate the next positions to punch the missing disks Tip You can easily toggle from the Plates view to the Punch view by bringing th...

Страница 17: ...utomatic identification system using running numbers This can be turned on or off from the Settings view for more details see Automatic Sample ID generation on page 24 To provide the sample ID manuall...

Страница 18: ...osition In Specimen Gate mode the right punch head is always brought to the front while waiting for the punch command as the instrument does not know the correct prioritized size before the queries ar...

Страница 19: ...ears red and each plate with a red well is outlined in red as well Select each plate outlined in red to see the detailed view Once you have visually checked the well you can change its status by selec...

Страница 20: ...ckground of the Punch head maintenance area also appears in red Even if the software reminds you to clean the punch heads you are be able to continue punching However it is strongly advised that you p...

Страница 21: ...emoving and cleaning the punch heads In order to clean a punch head you need to remove it from the puncher dismantle it and then clean the various parts The inner chute is the part that gathers the mo...

Страница 22: ...tor 90 degrees counterclockwise and slide it out The disk detector the chute and the spring come out of the punch head Remove the chute and the spring by pulling them Panthera Puncher 9 User Manual Ma...

Страница 23: ...Separate the spring from the chute In principle you should have the different components as shown in the picture below Panthera Puncher 9 User Manual Maintenance and shutdown 19...

Страница 24: ...ater first should largely eliminate this problem You may use swabs foam or nylon applicator sticks are best as cotton tipped applicator sticks can leave behind cotton filaments or cleaning clothes tha...

Страница 25: ...at a 90 degrees angle then insert the inner chute part into the guide and make sure that the fastening pins of the frame arm are placed in the groove of the plastic part of the detector You can then...

Страница 26: ...position and replace the punch head into the instrument Note On occasion when inserting the punch head back to the instrument after the daily punch head maintenance the Panthera software presents a Sk...

Страница 27: ...It is critical that you protect the detector from any liquid Panthera Puncher 9 User Manual Maintenance and shutdown 23...

Страница 28: ...0531 yy M d 11 5 31 MM dd BB 05 31BB AB yy ooo where ooo represents an ordinal date i e the number corresponding to the current day in the current year the 31st of May is the 151st day of the year 201...

Страница 29: ...ap autoincrement and start location parameters After pressing the button a new window opens and gives you the possibility to select a platemap to preview the platemap to set the start location for tha...

Страница 30: ...for a particular analyte press the corresponding button A page displays the analyte settings which are organized in six areas Punching Punches per well Card barcode rules Plate Plate maps Exporting Pu...

Страница 31: ...the rules for automatic identification of an analyte from a plate barcode See Card barcode rules on page 27 for basic instructions for writing rules For example if you want a rule for a plate barcode...

Страница 32: ...plate maps for the selected analyte Only three plate maps are previewed at a time You can browse through all plate maps by pressing the left and right arrows Every analyte must have at least one plat...

Страница 33: ...here x is the number of characters All static components in the barcode rules are written without The rules are case sensitive on the static components Table 1 Examples of barcode notation Rule Notati...

Страница 34: ...the sample cards punched on the corresponding day The name of each file is structured as follows xxxxxxxxx date time where xxxxxxxxx is of variable length and represents the sample card ID This is to...

Страница 35: ...are stored in the folder C ProgramData PerkinElmer SpotLight Logs The programdata folder is a hidden folder It can be accessed by typing the location in the address bar of Windows explorer Panthera Pu...

Страница 36: ...one more disk be punched from that same spot Is this normal This is due to the inaccuracy of pixel size calibration which ensures that the disks fit In some situations a disk can be a pixel smaller th...

Страница 37: ...about this product or you need assistance please contact PerkinElmer Customer Support Rest of the world GS customersupport perkinelmer com North America gs technical support perkinelmer com Panthera...

Страница 38: ...number listed below Give the full identification number of the manual you require This number is at the top right hand corner of the first page of the PDF The format is 2081 9xxx xx Each language ver...

Страница 39: ...ation 16 Maintenance and shutdown 16 O Operation Loading and setting up plates 9 Plate edit view 10 Punching and sample queries 12 Startup and shutdown 8 Operation continued Uloading plates from instr...

Страница 40: ...36...

Страница 41: ......

Страница 42: ...2081 9020 13 April 26 2022 Manufacturer Wallac Oy Mustionkatu 6 FI 20750 Turku Finland Tel 358 2 2678 111 Fax 358 2 2678 357 www perkinelmer com...

Отзывы: