Figure 9: Connecting tubing
To lock the connectors, insert the male connector into the female until it clicks. Used connectors are non-spill type (to
allow disconnection without leaking).
Figure 10: Tubing for the water bath
For efficient cooling, it is vital to remove any air from the cooling plate. Switch on and leave the circulator running until
all the air in the connecting tubing is removed, and this may take several minutes. Then refill the circulator reservoir
with cooling liquid (water)
Caution:
Ensure that tubing is not bent too severely (kinked). This blocks the coolant flow, and gel run may be
affected.
Note:
Using HI setting in the circulator pump speed is recommended.
Levelling the instrument
When the Migele Gel Electrophoresis Unit is in the final bench location, place a spirit level on the cooling plate (see
Figure 11: Spirit level to monitor that the instrument is levelled correctly
on page 15) (in inward position) and adjust
the unit foot until the unit is levelled (see
Figure 12: Adjust the unit with the levelling feet
2118-0010 - Migele
®
Gel Electrophoresis Unit User Manual
Installation
14