Setting the lid
Set the anode electrode positions according to your gel electrophoresis requirements. Use thumb screws to set the
position.
Connecting the power supply
Warning:
The supplied DC cables are rated for 5000 V. Only use cables and adapters supplied with the Migele Gel
Electrophoresis Unit or ensure that they have a suitable DC insulation compliance for the used voltages.
Connect the supplied wires from the power supply to the back plate of the Migele Gel Electrophoresis Unit (please use
wires of corresponding colors). One power supply can be used with two to four units in parallel depending on power
supply type used. See the pictures below for more details.
2118-0010 - Migele
®
Gel Electrophoresis Unit User Manual
Installation
17