38
FAN LED is lit on with an alarm sound
1.
Fan needs to be replaced.
LED of LOW LEVEL, FLOW SENSOR and PUMP blinks with an alarm sound
1.
Tunn off the power of the peltier temperature controller 201.
2.
Remove Phillips screws (4 each) on the Air Vent Manual button cover using Phillips screwdriver.
3.
Push the Air Vent Manual button on.
4.
Restart the Peltier Temperature Controller 201 and check the coolant level and if it is lower the ‘L’
(low) mark, fill up the coolant more.