09931463A
13
8.
If the coolant falls below ‘L’ (Low level), the LEDs of LOW LEVEL and FLOW SENSOR are lit on.
Figure 22 LEDs of LOW LEVEL and FLOW SENSOR
9.
When drain is completed, disconnect the waste hose.
10.
Turn off the AC power switch of the Peltier Temperature Controller 201.
11.
Push the Air Vent Manual button off.
12.
Fasten Phillips screws (4 each) on the Air Vent Manual button cover using Phillips screwdriver.
Installation
1.
Prepare the Lambda 365 and Peltier Temperature Controller 201 in a location that is compatible
with the required environmental conditions for the operation.
2.
Connect the power cord and communication cable of the Lambda 365.
DO NOT
turn on the
power of the instrument!
3.
Remove the two Phillips round head screws with washer (M4*12L) to disassemble the existing cell
holder and base plate.
Figure 23 Location of Phillips round head screws with washer
CAUTION
ATTENTION
Make sure the instrument is turned off while installing this accessory.
Assurez-vous que l’instrument est éteint lors de l’installation de cet accessoire.