09931463A
23
Measurement
Thermal Denaturation Mode
1.
Install the 6-Position Peltier Controlled Cell Changer referring to section
Installation
(page
13).
2.
Close the sample compartment cover and turn on the instrument.
NOTE:
Before System Self Test is performed, the 6-Position Peltier Controlled Cell Changer should be installed.
Otherwise, the instrument can be damaged electrically and do not operate properly.
NOTE:
Start the System Self Test after warming up the system for at least 20 minutes.
3.
Double-click on the
UV Express
folder and select
Thermal Denaturation
mode.
4.
Select
OK
after finishing the
System Self Test
.
5.
Select
New
to open a new window and select
OK
.
NOTE:
When the 6-Position Peltier Controlled Cell Changer is connected for the first time, you need to correct
the cell position. Calibrate the beam position of the 6-Position Peltier Controlled Cell Changer, see
section
Calibration of Multi-Cell Position
(page 33).
6.
Open the method in main menu or mode icon on the left side of the main window. Set parameters
for
Experiment
.