Troubleshooting
Power LED is not lit on
1.
Check the connection of the power cord or the fuse. The fuse is located at the rear of the
instrument.
WARNING
AVERTISSEMENT
There is risk of receiving a fatal electric shock if the fuses are replaced
with the power cord connected.
Il y a un risque d’électrocution si les fusibles sont remplacés tandis que
le cordon d'alimentation est encore branché.
2.
Turn off and unplug the instrument.
3.
Locate the fuse cover on the left rear panel of the instrument.
4.
Carefully open the compartment latch where the fuse is located.
5.
Disconnect the fuse.
6.
Replace with a new T5AL fuse (AC 250V). One spare is contained in the power module.
7.
Close the compartment door.
8.
Plug in the instrument and turn on.
On-Line LED is not lit on
1.
Check whether the communication cable is connected tightly.
2.
Change the port setting see section
Setting USB Serial Port
(page 19).
Location of latch on the
fuse compartment door