Appendices . 89
Initialization errors
If your instrument generates one of the following error messages, switch off the instrument,
then switch the instrument on again. Consult your PerkinElmer Service Engineer if the
problem persists.
Error message
Action
Program Integrity Failed
The instrument’s firmware may be corrupt, so must be
reinstalled.
Instrument Configuration
Failed
Suggests that one or more internal devices is failing. If
the problem persists, note any associated error
messages and consult your PerkinElmer service
representative.
Scan Failed
If the problem persists, note any associated error
messages and consult your PerkinElmer service
representative.
Laser Failed
Replace Laser
Replacing the Laser and Power Supply
page 83.
Source Failed
Replace Source
on page 75.
Calibration Failed
Too little energy may be reaching the detector. Make
sure the sample compartment is clear.
Содержание Frontier Optica FT-IR
Страница 1: ...FRONTIER OPTICA FT IR User s Guide MOLECULAR SPECTROSCOPY ...
Страница 5: ...Introduction ...
Страница 10: ...10 Frontier Optica FT IR User s Guide ...
Страница 11: ...Warnings and Safety Information ...
Страница 23: ...An Overview of the Frontier Optica FT IR Spectrometer ...
Страница 29: ...Unpacking and Installation ...
Страница 40: ...40 Frontier Optica FT IR User s Guide ...
Страница 41: ...Using the Spectrometer ...
Страница 50: ...50 Frontier Optica FT IR User s Guide ...
Страница 51: ...Measuring Optical Components ...
Страница 58: ...58 Frontier Optica FT IR User s Guide ...
Страница 59: ...Routine Maintenance ...
Страница 71: ...Advanced Maintenance ...
Страница 85: ...Appendices ...