Advanced Maintenance . 79
Installing Filters in the Filter Wheel
The following procedure describes how to install a filter in the filter wheel. The same
procedure should be followed when replacing a filter or the methane cell.
The filter wheel assembly is removed from its mounting and placed on the standard sample
slide baseplate.
1.
Set the
Filter Wheel
to the Position 1 on the Setup Instrument BeamPath tab in the
Spectrum software.
Position 1 will be labelled as
None 15000 – 0 cm
-1
and be the first option in the drop-
down list.
WARNING
Switch off the mains power supply to the spectrometer, wait 60 seconds,
and disconnect the power cable before you open the cover of the
spectrometer. This makes sure that you are safe from electrical shock
and laser radiation.
2.
Open the main cover of the instrument as detailed on page 72 and identify the filter
wheel.
Figure 42 Filter wheel location
Filter wheel
Содержание Frontier Optica FT-IR
Страница 1: ...FRONTIER OPTICA FT IR User s Guide MOLECULAR SPECTROSCOPY ...
Страница 5: ...Introduction ...
Страница 10: ...10 Frontier Optica FT IR User s Guide ...
Страница 11: ...Warnings and Safety Information ...
Страница 23: ...An Overview of the Frontier Optica FT IR Spectrometer ...
Страница 29: ...Unpacking and Installation ...
Страница 40: ...40 Frontier Optica FT IR User s Guide ...
Страница 41: ...Using the Spectrometer ...
Страница 50: ...50 Frontier Optica FT IR User s Guide ...
Страница 51: ...Measuring Optical Components ...
Страница 58: ...58 Frontier Optica FT IR User s Guide ...
Страница 59: ...Routine Maintenance ...
Страница 71: ...Advanced Maintenance ...
Страница 85: ...Appendices ...