36 . Frontier Optica FT-IR User's Guide
Other connectors
The communication ports for peripheral devices are shown in Figure 10.
Figure 10 Communication ports
Icons that identify the function of each of the Communication ports are printed on the hinge
molding directly above the port.
Connector details
Icon
Description
Connector
Type
Voltages Maximum
Currents
10101
Serial port. Connects a PC to the
instrument using an RS232
interface cable. This connector is
only used for service diagnostics.
9-way
D-type
+12 V
<100 mA
−
12 V
<100 mA
General Purpose Optical Bench
(GPOB). Allows synchronization
between the instrument and an
external accessory.
26-way
high density
D-type
+5 V
<50 mA
Spt. Lt
Spotlight. Used to connect the
instrument to a Spotlight Imaging
System.
11-way
mixed
D-type
+5 V
<100 mA
±
12 V
50 mA
Sample area purge
Instrument purge
10101
Spt. Lt.
EXT.R
Содержание Frontier Optica FT-IR
Страница 1: ...FRONTIER OPTICA FT IR User s Guide MOLECULAR SPECTROSCOPY ...
Страница 5: ...Introduction ...
Страница 10: ...10 Frontier Optica FT IR User s Guide ...
Страница 11: ...Warnings and Safety Information ...
Страница 23: ...An Overview of the Frontier Optica FT IR Spectrometer ...
Страница 29: ...Unpacking and Installation ...
Страница 40: ...40 Frontier Optica FT IR User s Guide ...
Страница 41: ...Using the Spectrometer ...
Страница 50: ...50 Frontier Optica FT IR User s Guide ...
Страница 51: ...Measuring Optical Components ...
Страница 58: ...58 Frontier Optica FT IR User s Guide ...
Страница 59: ...Routine Maintenance ...
Страница 71: ...Advanced Maintenance ...
Страница 85: ...Appendices ...