66 . Frontier IR Single-range Spectrometers User's Guide
1.
Remove the MCT coolant cover (Figure 34).
Figure 34 Location of MCT coolant cover
2.
Remove the plug from the dewar.
3.
Ensure that the funnel guide, which helps prevent overspill entering the instrument
body rather than the dewar, is in place.
4.
Make sure the supplied plastic funnel is dry.
CAUTION
Stand where you can see the inside of the funnel as you pour the
nitrogen in. Pour slowly, so that neither the funnel nor the dewar
overflows. Take care not to overfill the funnel and splash liquid nitrogen
onto the instrument covers.
5.
Insert the plastic funnel into the funnel guide and slowly fill the detector dewar with
liquid nitrogen.
NOTE:
Make sure that the detector dewar is filled to the top with liquid nitrogen before
selecting the detector. Prevent radiation from reaching the detector by obstructing
the beam path in the sample area during filling.
6.
Stand back and wait two minutes.
This nitrogen also vaporizes as the dewar continues to cool. The two-minute wait allows
the bubbling to settle down and the pressure of the vaporizing nitrogen to dissipate.
7.
Continue to pour liquid nitrogen into the funnel, adding a little more each time the
funnel empties.
The funnel takes longer to empty as the dewar fills. This happens after two or three
more funnels of liquid nitrogen. Because the dewar has now cooled, the liquid nitrogen
does not vaporize, but instead fills the dewar.
8.
Remove the funnel and wait two minutes.
The liquid nitrogen settles down and bubbling slows.
MCT coolant cover
Содержание Frontier FT-IR
Страница 1: ...FRONTIER IR SINGLE RANGE SYSTEMS User s Guide MOLECULAR SPECTROSCOPY ...
Страница 5: ...Introduction ...
Страница 10: ...10 Frontier IR Single range Spectrometers User s Guide ...
Страница 11: ...Warnings and Safety Information ...
Страница 23: ...An Overview of the Frontier IR Single Range Spectrometers ...
Страница 32: ...32 Frontier IR Single range Spectrometers User s Guide ...
Страница 33: ...Unpacking and Installation ...
Страница 43: ...Using the Spectrometer with Spectrum ...
Страница 49: ...Using the Spectrometer with Spectrum 49 Figure 20 Atmospheric suppression ...
Страница 53: ...Routine Maintenance ...
Страница 68: ...68 Frontier IR Single range Spectrometers User s Guide ...
Страница 69: ...Advanced Maintenance ...
Страница 88: ...88 Frontier IR Single range Spectrometers User s Guide ...
Страница 89: ...Appendices ...
Страница 100: ...100 Frontier IR Single range Spectrometers User s Guide ...