74 . Frontier IR Single-range Spectrometers User's Guide
Replacing the Source
WARNING
Switch off the mains power supply to the spectrometer, wait
60 seconds and disconnect the power cable before you open the cover of
the spectrometer. This makes sure that you are safe from electrical shock
and laser radiation.
WARNING
Due to its high operating temperature allow sufficient time for the source
to cool before attempting to remove it.
1.
Open the main cover of the instrument as detailed on page 70 and identify the source
Figure 41 Source location
Caution, hot surface.
The source housing and source itself will be hot if recently switched on.
Allow to cool before performing the following procedure.
CAUTION
The source element is brittle and can be broken if not handled properly.
Be careful when removing and installing the source.
Do not touch the glass bulb as this will degrade the performance of the
bulb and shorten its life.
Source
Содержание Frontier FT-IR
Страница 1: ...FRONTIER IR SINGLE RANGE SYSTEMS User s Guide MOLECULAR SPECTROSCOPY ...
Страница 5: ...Introduction ...
Страница 10: ...10 Frontier IR Single range Spectrometers User s Guide ...
Страница 11: ...Warnings and Safety Information ...
Страница 23: ...An Overview of the Frontier IR Single Range Spectrometers ...
Страница 32: ...32 Frontier IR Single range Spectrometers User s Guide ...
Страница 33: ...Unpacking and Installation ...
Страница 43: ...Using the Spectrometer with Spectrum ...
Страница 49: ...Using the Spectrometer with Spectrum 49 Figure 20 Atmospheric suppression ...
Страница 53: ...Routine Maintenance ...
Страница 68: ...68 Frontier IR Single range Spectrometers User s Guide ...
Страница 69: ...Advanced Maintenance ...
Страница 88: ...88 Frontier IR Single range Spectrometers User s Guide ...
Страница 89: ...Appendices ...
Страница 100: ...100 Frontier IR Single range Spectrometers User s Guide ...