76 . Frontier IR Single-range Spectrometers User's Guide
Replacing the Beamsplitter
WARNING
Switch off the mains power supply to the spectrometer, wait 60 seconds,
and disconnect the power cable before you open the cover of the
spectrometer. This makes sure that you are safe from electrical shock
and laser radiation.
1.
Open the main cover of the instrument, as detailed on page 70 and identify the
Beamsplitter, which is behind the laser baffle (Figure 43).
Figure 43 Beamsplitter location
2.
Unscrew the laser baffle retaining screws and remove the laser baffle (Figure 44).
Figure 44 Laser baffle
3.
Remove the beamsplitter retaining screw.
4.
Noting its installed position, pull out the beamsplitter.
5.
Install the new beamsplitter and secure with the retaining screw.
Laser baffle
retaining screw
(silver)
Laser baffle
retaining screw
(silver)
Beamsplitter
retaining screw
(black)
Beamsplitter
(behind baffle)
Laser baffle
Содержание Frontier FT-IR
Страница 1: ...FRONTIER IR SINGLE RANGE SYSTEMS User s Guide MOLECULAR SPECTROSCOPY ...
Страница 5: ...Introduction ...
Страница 10: ...10 Frontier IR Single range Spectrometers User s Guide ...
Страница 11: ...Warnings and Safety Information ...
Страница 23: ...An Overview of the Frontier IR Single Range Spectrometers ...
Страница 32: ...32 Frontier IR Single range Spectrometers User s Guide ...
Страница 33: ...Unpacking and Installation ...
Страница 43: ...Using the Spectrometer with Spectrum ...
Страница 49: ...Using the Spectrometer with Spectrum 49 Figure 20 Atmospheric suppression ...
Страница 53: ...Routine Maintenance ...
Страница 68: ...68 Frontier IR Single range Spectrometers User s Guide ...
Страница 69: ...Advanced Maintenance ...
Страница 88: ...88 Frontier IR Single range Spectrometers User s Guide ...
Страница 89: ...Appendices ...
Страница 100: ...100 Frontier IR Single range Spectrometers User s Guide ...