Routine Maintenance . 57
Changing the Desiccant
CAUTION
Expect to change the desiccant
in the
spectrometer every six months.
Old, used desiccant releases moisture.
In regions experiencing high humidity levels we recommend that you
change the desiccant more often.
The desiccant change interval is set in the
Desiccant change due in (days)
option on the
Setup Instrument BeamPath tab in Spectrum software (Figure 24).
Figure 24 Setup Instrument BeamPath tab in Spectrum
If an instrument desiccant change is overdue, a warning will be displayed in the Status Bar
and in the Setup Instrument BeamPath table.
The warning is displayed until you tell the software that the desiccant has been changed. To
reset the desiccant change interval, select the current value, click
Changed
to clear any
warning messages, and then enter the number of days before the next desiccant change is
due.
Содержание Frontier FT-IR
Страница 1: ...FRONTIER IR SINGLE RANGE SYSTEMS User s Guide MOLECULAR SPECTROSCOPY ...
Страница 5: ...Introduction ...
Страница 10: ...10 Frontier IR Single range Spectrometers User s Guide ...
Страница 11: ...Warnings and Safety Information ...
Страница 23: ...An Overview of the Frontier IR Single Range Spectrometers ...
Страница 32: ...32 Frontier IR Single range Spectrometers User s Guide ...
Страница 33: ...Unpacking and Installation ...
Страница 43: ...Using the Spectrometer with Spectrum ...
Страница 49: ...Using the Spectrometer with Spectrum 49 Figure 20 Atmospheric suppression ...
Страница 53: ...Routine Maintenance ...
Страница 68: ...68 Frontier IR Single range Spectrometers User s Guide ...
Страница 69: ...Advanced Maintenance ...
Страница 88: ...88 Frontier IR Single range Spectrometers User s Guide ...
Страница 89: ...Appendices ...
Страница 100: ...100 Frontier IR Single range Spectrometers User s Guide ...