COMMON ERRORS FOUND WITH ELECTROPHORESIS POWER SUPPLIES 

No load 

 

The electrophoresis system is not connected to the power leads, check the power leads  

 

The electrophoresis system has a short, the Pt wire is broken or the banana connectors are dam-
aged 

 

Buffer concentration too low 

 

Buffer volume too low 

 

Short in power cord 

 

Current has dropped below acceptable rating (4 mA) 

Short circuit 

 

Load exceeds 500 mA 

 

Blown fuse in the power supply 

 

Incorrect input voltage (check input voltage switch near power inlet) 

Change in load 

 

Electrophoresis systems were added or removed during a run 

 

Buffer leaking in a connected system 

 

Excessive temperature increase 

 

Excessive buffer evaporation 

 

Loose connection in a connected system 

 

Amperage set to low 

Change in constant mode 

 

Voltage changes to amperage 

o

 

Amperage set too low. Ceiling hit and constant mode changed from voltage to amper-
age. Increase amperage to 500 mA. 

 

Amperage changes to voltage 

o

 

Voltage set too low. Ceiling hit and constant mode changed from amperage to voltage. 
Increase voltage to 300 volts 

 
The 300 volt system has automatic cross over, set voltage or amperage, and preset wattage. During the 
electrophoresis process only one parameter is limiting at a time. The limiting parameter, together with 
the conductivity in the electrophoresis system, and the values for the other parameters determine the 
maximum output. 

PEQLAB_v0412_E 

12

Содержание peqPOWER 300

Страница 1: ...Instruction Manual 300 Volt Power Supply 0412E Creating the future together...

Страница 2: ......

Страница 3: ...RTED 6 IMPORTANT GUIDELINES 7 OPERATIONAL MODES 8 CHOOSING LIMITING PARAMETER SETTINGS 9 TROUBLESHOOTING 10 OPERATIONAL ELECTRICAL PARAMETERS 11 COMMON ERRORS FOUND WITH ELECTROPHORESIS POWER SUPPLIES...

Страница 4: ...shielded banana plugs thus minimizing any potential shock hazard to the user Always use gel box sys tems that are compatible with the Power Supply have been designed for your specific applications an...

Страница 5: ...ease check carefully that the shipment is complete and has not been dam aged in transit For missing parts or to report any kind of damage please contact PEQLAB see TECH NICAL SERVICE AND ORDERING INFO...

Страница 6: ...age and 4 negative voltage Operating Modes Constant Voltage 1 V step Constant Current 1 mA step Crossover Automatic Display type Backlit LCD Graphic type Display size 53 64 x 15 64 mm W x H Pause func...

Страница 7: ...f lab space The peqPOWER 300 Volt Power Supply offers two modes Constant Voltage or Constant Current Mode This manual describes the setup and operation of peqPOWER 300 Volt Power Supply including impo...

Страница 8: ...Supply PEQLAB_v0412_E 5 Description of Buttons and Switches Front View of peqPOWER 300 Volt Power Supply Rear View of peqPOWER 300 Volt Power Supply Control Buttons LCD Display Output Jacks Voltage Se...

Страница 9: ...nu meric values STOP START PAUSE CONSTANT MODE GETTING STARTED Installing the peqPOWER 300 Volt Power Supply 1 Check the label located near the AC inlet to ensure that the unit is compatible with loc...

Страница 10: ...the peqPOWER 300 Volt Power Supply by pressing the power switch the rear side of the instrument Upon start up the Display Screen on the front of the instrument will illuminate Use the START PAUSE and...

Страница 11: ...nual before starting electrophoresis using the peqPOWER 300 Volt Power Supply A basic Constant Voltage Current Mode operating procedure of the peqPOWER 300 Volt Power Sup ply is provided below We reco...

Страница 12: ...out the run providing a greater margin of safety over time The same voltage setting can be used regardless of the number or thickness of gels being proc essed Current Limiting Discontinuous buffer sys...

Страница 13: ...electrophoresis cell 5 Make sure buffers are prepared correctly and the recommended volume of buffer is added to the electrophoresis unit Operation stops with alarm The screen displays NO LOAD 6 High...

Страница 14: ...n Although the resistance is determined by the gel system the resistance can vary over the course of an electrophoretic separation Voltage The velocity with which an ion moves in an electric field wil...

Страница 15: ...uring a run Buffer leaking in a connected system Excessive temperature increase Excessive buffer evaporation Loose connection in a connected system Amperage set to low Change in constant mode Voltage...

Страница 16: ...arts are dried thoroughly before attempting to operate the unit Do not immerse the centrifuge in liquid or pour liquids over it Before using any cleaning or decontamination methods except those recomm...

Страница 17: ...ent Power Cat No peqPOWER 250 V 230 V 4 250 3000 mA 300 W 55 E250 230V peqPOWER 250 V 110 V 4 250 3000 mA 300 W 55 E250 110V For technical questions and more detailed information on PEQLAB s products...

Страница 18: ...WARRANTY OF MERCHANTABILITY For research use only Not intended for any animal or human therapeutic or diagnostic use EQUIPMENT DISPOSAL 15 This equipment is marked with the crossed out wheeled bin sym...

Страница 19: ......

Страница 20: ...QLAB Biotechnologie GmbH 6404 Polling Tel 43 0 5238 84 169 info peqlab at www peqlab at UK PEQLAB Ltd Southampton SO31 7ZN Freecall UK 0808 20 21 302 info peqlab co uk www peqlab co uk USA PEQLAB LLC...

Отзывы: