Instruction Manual peqPOWER 300 Volt Power Supply
COMMON ERRORS FOUND WITH ELECTROPHORESIS POWER SUPPLIES
No load
The electrophoresis system is not connected to the power leads, check the power leads
The electrophoresis system has a short, the Pt wire is broken or the banana connectors are dam-
aged
Buffer concentration too low
Buffer volume too low
Short in power cord
Current has dropped below acceptable rating (4 mA)
Short circuit
Load exceeds 500 mA
Blown fuse in the power supply
Incorrect input voltage (check input voltage switch near power inlet)
Change in load
Electrophoresis systems were added or removed during a run
Buffer leaking in a connected system
Excessive temperature increase
Excessive buffer evaporation
Loose connection in a connected system
Amperage set to low
Change in constant mode
Voltage changes to amperage
o
Amperage set too low. Ceiling hit and constant mode changed from voltage to amper-
age. Increase amperage to 500 mA.
Amperage changes to voltage
o
Voltage set too low. Ceiling hit and constant mode changed from amperage to voltage.
Increase voltage to 300 volts
The 300 volt system has automatic cross over, set voltage or amperage, and preset wattage. During the
electrophoresis process only one parameter is limiting at a time. The limiting parameter, together with
the conductivity in the electrophoresis system, and the values for the other parameters determine the
maximum output.