OPERATIONAL MODES 

Introduction 

The peqPOWER 300 Volt Power Supply is designed to operate under two modes, Constant Voltage 
mode or Constant Current mode, depending upon your electrophoresis application. 
Use the Constant Voltage / Constant Current Mode for applications that require only one specific volt-
age limit or current limit during the entire duration of electrophoresis. 

Display Screen 

The Display Screen illuminates after turning on the power to your instrument and the factory default 
settings (or last settings) will be displayed. You can choose the operational Mode (Constant Voltage or 
Constant Current Mode) by pressing the “constant” button. 
On the Display Screen: 

-

 

The chosen constant parameter (Voltage or Current) is displayed in bold on the left 
side of the display. 

-

 

The Timer is the first line on the right-top, and the non-constant value is displayed in 
the second line on the right side of the display screen. 

Basic Constant Operation Protocol 

The Constant Voltage and Constant Current Modes allow you to specify a voltage limit, and current limit 
to be used continuously during the entire duration of electrophoresis. Review the guidelines provided in 
this manual before starting electrophoresis using the peqPOWER 300 Volt Power Supply. 
A basic Constant Voltage / Current Mode operating procedure of the peqPOWER 300 Volt Power Sup-
ply is provided below. We recommend reading the guidelines provided in this manual for best results 
before starting an operation. 

1.

 

Use the power switch on the rear side of the instrument to turn on the peqPOWER 300 Volt 
Power Supply. The Display Screen will illuminate. 

2.

 

Press the CONSTANT key to select either Constant Voltage Operation or Constant Current Op-
eration from the Display Screen. 

3.

 

Use the Up arrow / Down arrow keys to set either voltage (V) or current (mA) parameter to the 
appropriate values. 

4.

 

Press the MODE key to choose the TIME parameter, and use the Up Arrow / Down Arrow keys 
to set the duration (hours/minutes) of the electrophoresis run. 

5.

 

Press START/PAUSE key to start electrophoresis. 

6.

 

Press the START/PAUSE key again to temporarily interrupt power. The red “Run”LED will flash 
to indicate that the electrophoresis run is paused. Pressing the START/PAUSE key again will re-
start the electrophoresis run. 

7.

 

Press the STOP key to permanently stop the electrophoresis run (the timer will reset). 

8.

 

To change the limits (Voltage or Amperage) of the electrophoresis run in progress: Press the 
Mode key. Enter the changes using the Up Arrow / Down Arrow keys, and then press 
START/PAUSE key once again to restart your operation. 

PEQLAB_v0412_E 

8

Содержание peqPOWER 300

Страница 1: ...Instruction Manual 300 Volt Power Supply 0412E Creating the future together...

Страница 2: ......

Страница 3: ...RTED 6 IMPORTANT GUIDELINES 7 OPERATIONAL MODES 8 CHOOSING LIMITING PARAMETER SETTINGS 9 TROUBLESHOOTING 10 OPERATIONAL ELECTRICAL PARAMETERS 11 COMMON ERRORS FOUND WITH ELECTROPHORESIS POWER SUPPLIES...

Страница 4: ...shielded banana plugs thus minimizing any potential shock hazard to the user Always use gel box sys tems that are compatible with the Power Supply have been designed for your specific applications an...

Страница 5: ...ease check carefully that the shipment is complete and has not been dam aged in transit For missing parts or to report any kind of damage please contact PEQLAB see TECH NICAL SERVICE AND ORDERING INFO...

Страница 6: ...age and 4 negative voltage Operating Modes Constant Voltage 1 V step Constant Current 1 mA step Crossover Automatic Display type Backlit LCD Graphic type Display size 53 64 x 15 64 mm W x H Pause func...

Страница 7: ...f lab space The peqPOWER 300 Volt Power Supply offers two modes Constant Voltage or Constant Current Mode This manual describes the setup and operation of peqPOWER 300 Volt Power Supply including impo...

Страница 8: ...Supply PEQLAB_v0412_E 5 Description of Buttons and Switches Front View of peqPOWER 300 Volt Power Supply Rear View of peqPOWER 300 Volt Power Supply Control Buttons LCD Display Output Jacks Voltage Se...

Страница 9: ...nu meric values STOP START PAUSE CONSTANT MODE GETTING STARTED Installing the peqPOWER 300 Volt Power Supply 1 Check the label located near the AC inlet to ensure that the unit is compatible with loc...

Страница 10: ...the peqPOWER 300 Volt Power Supply by pressing the power switch the rear side of the instrument Upon start up the Display Screen on the front of the instrument will illuminate Use the START PAUSE and...

Страница 11: ...nual before starting electrophoresis using the peqPOWER 300 Volt Power Supply A basic Constant Voltage Current Mode operating procedure of the peqPOWER 300 Volt Power Sup ply is provided below We reco...

Страница 12: ...out the run providing a greater margin of safety over time The same voltage setting can be used regardless of the number or thickness of gels being proc essed Current Limiting Discontinuous buffer sys...

Страница 13: ...electrophoresis cell 5 Make sure buffers are prepared correctly and the recommended volume of buffer is added to the electrophoresis unit Operation stops with alarm The screen displays NO LOAD 6 High...

Страница 14: ...n Although the resistance is determined by the gel system the resistance can vary over the course of an electrophoretic separation Voltage The velocity with which an ion moves in an electric field wil...

Страница 15: ...uring a run Buffer leaking in a connected system Excessive temperature increase Excessive buffer evaporation Loose connection in a connected system Amperage set to low Change in constant mode Voltage...

Страница 16: ...arts are dried thoroughly before attempting to operate the unit Do not immerse the centrifuge in liquid or pour liquids over it Before using any cleaning or decontamination methods except those recomm...

Страница 17: ...ent Power Cat No peqPOWER 250 V 230 V 4 250 3000 mA 300 W 55 E250 230V peqPOWER 250 V 110 V 4 250 3000 mA 300 W 55 E250 110V For technical questions and more detailed information on PEQLAB s products...

Страница 18: ...WARRANTY OF MERCHANTABILITY For research use only Not intended for any animal or human therapeutic or diagnostic use EQUIPMENT DISPOSAL 15 This equipment is marked with the crossed out wheeled bin sym...

Страница 19: ......

Страница 20: ...QLAB Biotechnologie GmbH 6404 Polling Tel 43 0 5238 84 169 info peqlab at www peqlab at UK PEQLAB Ltd Southampton SO31 7ZN Freecall UK 0808 20 21 302 info peqlab co uk www peqlab co uk USA PEQLAB LLC...

Отзывы: