TROUBLESHOOTING 

Review the information in the table below to troubleshoot operating problems. 

Problem Cause 

Solution 

AC power cord is not connected  Check AC power cord connections at both 

ends. Use the correct cords. 

The LCD screen 
remains blank and 
the fan does not run 
when the power is 
turned on 

The fuse has blown 

Replace the fuse 

1. Electrophoresis leads are not 
connected to the power supply. 

1. Check the connections to the power supply 
and on your electrophoresis cell to make sure 
the connection is intact; check condition of 
wires in electrophoresis unit.  

2. The electrophoresis unit(s), is 
not connected. 

2. Close the circuit by reconnecting the ca-
bles.  

3. There is a broken circuit in 
the electrophoresis cell 

3. Press START/PAUSE to restart the run. 

4. High resistance due to tape 
left on a pre-cast gel  

4. Correct the condition by making sure the 
tape is removed from the pre-cast gel. 

5. Incorrect buffer concentra-
tion, or incorrect buffer volumes 
in the electrophoresis cell 

5. Make sure buffers are prepared correctly, 
and the recommended volume of buffer is 
added to the electrophoresis unit. 

Operation stops 
with alarm 
The screen displays 
“NO LOAD” 

6. High voltage application is 
set to run on a very low current 

6. Change current setting. 

Operation stops 
with alarm: 
Display shows 
“OVER VOLTAGE” 

Circuit is interrupted 
 

1. Verify that the running buffer is correct. 
2. Verify the all cables are attached correctly 
3. Turn the Power switch off and on again; 
restart application. 
4. If you cannot restart the instrument, turn off 
the power, disconnect the power cord from 
the outlet, and contact Technical Service. 

Operation stops 
with alarm: 
Display shows 
“LEAKAGE” 

Ground leak detected during 
run 

1. Check the electrophoresis system for im-
proper grounding.  
2. Restart the power supply by turning the 
Power switch off and on. 

Operation stops 
with alarm: 
Display shows 
“OVER TEMP” 

Power supply is overheating 

1. Turn off power supply. Check for sufficient 
airflow around the power supply fan. After 
cooling down, restart the power supply. 
2. If you cannot restart the instrument, turn off 
the power, disconnect the power cord from 
the outlet, and contact Technical Service. 

 
 

PEQLAB_v0412_E 

10

Содержание peqPOWER 300

Страница 1: ...Instruction Manual 300 Volt Power Supply 0412E Creating the future together...

Страница 2: ......

Страница 3: ...RTED 6 IMPORTANT GUIDELINES 7 OPERATIONAL MODES 8 CHOOSING LIMITING PARAMETER SETTINGS 9 TROUBLESHOOTING 10 OPERATIONAL ELECTRICAL PARAMETERS 11 COMMON ERRORS FOUND WITH ELECTROPHORESIS POWER SUPPLIES...

Страница 4: ...shielded banana plugs thus minimizing any potential shock hazard to the user Always use gel box sys tems that are compatible with the Power Supply have been designed for your specific applications an...

Страница 5: ...ease check carefully that the shipment is complete and has not been dam aged in transit For missing parts or to report any kind of damage please contact PEQLAB see TECH NICAL SERVICE AND ORDERING INFO...

Страница 6: ...age and 4 negative voltage Operating Modes Constant Voltage 1 V step Constant Current 1 mA step Crossover Automatic Display type Backlit LCD Graphic type Display size 53 64 x 15 64 mm W x H Pause func...

Страница 7: ...f lab space The peqPOWER 300 Volt Power Supply offers two modes Constant Voltage or Constant Current Mode This manual describes the setup and operation of peqPOWER 300 Volt Power Supply including impo...

Страница 8: ...Supply PEQLAB_v0412_E 5 Description of Buttons and Switches Front View of peqPOWER 300 Volt Power Supply Rear View of peqPOWER 300 Volt Power Supply Control Buttons LCD Display Output Jacks Voltage Se...

Страница 9: ...nu meric values STOP START PAUSE CONSTANT MODE GETTING STARTED Installing the peqPOWER 300 Volt Power Supply 1 Check the label located near the AC inlet to ensure that the unit is compatible with loc...

Страница 10: ...the peqPOWER 300 Volt Power Supply by pressing the power switch the rear side of the instrument Upon start up the Display Screen on the front of the instrument will illuminate Use the START PAUSE and...

Страница 11: ...nual before starting electrophoresis using the peqPOWER 300 Volt Power Supply A basic Constant Voltage Current Mode operating procedure of the peqPOWER 300 Volt Power Sup ply is provided below We reco...

Страница 12: ...out the run providing a greater margin of safety over time The same voltage setting can be used regardless of the number or thickness of gels being proc essed Current Limiting Discontinuous buffer sys...

Страница 13: ...electrophoresis cell 5 Make sure buffers are prepared correctly and the recommended volume of buffer is added to the electrophoresis unit Operation stops with alarm The screen displays NO LOAD 6 High...

Страница 14: ...n Although the resistance is determined by the gel system the resistance can vary over the course of an electrophoretic separation Voltage The velocity with which an ion moves in an electric field wil...

Страница 15: ...uring a run Buffer leaking in a connected system Excessive temperature increase Excessive buffer evaporation Loose connection in a connected system Amperage set to low Change in constant mode Voltage...

Страница 16: ...arts are dried thoroughly before attempting to operate the unit Do not immerse the centrifuge in liquid or pour liquids over it Before using any cleaning or decontamination methods except those recomm...

Страница 17: ...ent Power Cat No peqPOWER 250 V 230 V 4 250 3000 mA 300 W 55 E250 230V peqPOWER 250 V 110 V 4 250 3000 mA 300 W 55 E250 110V For technical questions and more detailed information on PEQLAB s products...

Страница 18: ...WARRANTY OF MERCHANTABILITY For research use only Not intended for any animal or human therapeutic or diagnostic use EQUIPMENT DISPOSAL 15 This equipment is marked with the crossed out wheeled bin sym...

Страница 19: ......

Страница 20: ...QLAB Biotechnologie GmbH 6404 Polling Tel 43 0 5238 84 169 info peqlab at www peqlab at UK PEQLAB Ltd Southampton SO31 7ZN Freecall UK 0808 20 21 302 info peqlab co uk www peqlab co uk USA PEQLAB LLC...

Отзывы: