EVERPURE

®

 CONSERV

®

 LT

-

S HIGH EFFICIENCY RO SYSTEM INSTALLATION AND OPERATION GUIDE

WARNING: Before installing the Processor, make 

certain your water supply complies with 

the following operating specifications. 

Failure to do so may reduce the 

effectiveness of the system and will void 

the warranty. Consult your local water 

treatment utility or a certified water 

testing lab to determine the quality of your 

water.

WARNING: Do not use with water that is 

microbiologically unsafe or of unknown 

quality without adequate disinfection 

before or after the system. The Everpure

®

 

Conserv

®

 LT-S Processor will not protect 

against disease-causing bacteria or 

remove harmless, naturally-occurring 

bacteria. 

WARNING: The Conserv LT-S Processor contains 

replaceable components that are critical to 

the efficiency of the system. Replacement 

of the reverse osmosis component should 

be with one of identical specifications, as 

defined by the manufacturer, to assure the 

same efficiency and contaminant reduction 

performance.

WARNING: The Conserv LT-S Processor contains filter 

cartridges and an RO membrane cartridge 

included with the system that have limited 

service lives. Product water should be 

tested periodically to verify that the 

Processor is working properly.

WARNING: The Conserv LT-S Processor shall only be 

used for arsenic reduction on chlorinated 

water supplies containing detectable 

residual free chlorine at the system inlet. 

Water systems using an in-line chlorinator 

should provide one (1) minute of chlorine 

contact time before the unit.

CAUTION:  The Conserv LT-S Processor must be 

protected against freezing which can 

cause the filter housing to crack, resulting 

in water leakage. Do not use electrical 

heating tape or propane torch on this unit.

CAUTION:  Make certain the unit is firmly attached to 

wall to prevent it from falling and possibly 

becoming damaged.

NOTE: Substances listed as reduced are not 

necessarily in your water. The Processor must 

be maintained according to manufacturer's 

instructions, including replacement of filter 

cartridges.

NOTE: Your water must be within required limits 

for satisfactory operation. If not, the RO 

membrane cartridge’s life may be shortened 

and your warranty will be voided (see 

Operating Specifications).

NOTE: Install on cold water line only. 

NOTE: Do not install where system will be exposed to 

direct sunlight. 

NOTE: Make certain that installation complies with all 

state and local laws and regulations. 

NOTE: Use only plumbers tape without adhesive 

backing to seal joints. Do not use pipe 

compound (“pipe dope”), sticks, or similar 

compounds with this unit; they contain 

petroleum derivatives which can cause 

crazing and cracking of the plastic in the filter 

housing.

NOTE: Use only soap and water to clean components.
NOTE: Do not use aerosol sprays (bug spray, cleaning 

fluids, etc.) near the Conserv LT-S Processor. 

They contain organic solvents which can cause 

crazing and cracking of the plastic in the filter 

housing.

NOTE: During extended periods of non-use, turn OFF 

the water supply to the unit and remove the 

cartridges. Place the cartridges in a sealed 

plastic bag. Store them in a refrigerator for 

future use. When restarting the unit, the 

cartridges are reinstalled and flushed per 

instructions.

OPERATING SPECIFICATIONS

WARNING: Before installing the Processor, make 

certain your water supply complies with 

the following operating specifications. 

Failure to do so may reduce the 

effectiveness of the system and will void 

the warranty. Consult your local water 

treatment utility or a certified water 

testing lab to determine the quality of  

your water.

Inlet Temperature:

32 - 100°F (0 - 37.7°C)

Dynamic Operating 

Pressure:

40 - 85 psi (2.7 - 5.8 bar, 0.27 - 0.58 

MPa), non-shock

Maximum Static 

Pressure

125 psi (8.6 bar, 0.86 MPa)

Inlet Total Dissolved 

Solids:

1,000 ppm maximum

Sulfide, Iron and 

Manganese:

 

Less than 0.1 ppm

Water Supply pH Limits:

4 to 11

Turbidity:

1.0 NTU maximum

For cold water use only.

Содержание EVERPURE CONSERV LT-S

Страница 1: ...ounting the Conserv LT S Processor inspect the parts Use the table and illustration provided to verify the parts provided If any parts are missing contact Customer Service at 800 942 1153 US only 630 307 3000 or by email at CSeverpure pentair com Parts 1 Wall Unit with GRO 75EN RO Membrane Cartridge in Place 2 4CC Mineral Cartridge 3 4FC5 Filter Cartridge 4 Storage Tank 5 Parts Bag Containing Bypa...

Страница 2: ...s restricted to trained adult personnel No one should play with the appliance NOTE Children should be supervised to ensure that they do not play with the appliance NOTE Cleaning and user maintenance shall not be made by children without supervision NOTE The appliance is not suitable for installation in an area where a water jet could be used NOTE If the supply cord is damaged it must be replaced b...

Страница 3: ...d NOTE Substances listed as reduced are not necessarily in your water The Processor must be maintained according to manufacturer s instructions including replacement of filter cartridges NOTE Your water must be within required limits for satisfactory operation If not the RO membrane cartridge s life may be shortened and your warranty will be voided see Operating Specifications NOTE Install on cold...

Страница 4: ...4 WATER YOU CAN TRUST EVERPURE CONSERV LT S PROCESSOR 1 Secure Unit to Wall Approximately 25 lbs 11 3 kgs Operating Weight 1 3 2 Remove RO Membrane 2 3 ...

Страница 5: ...ge Tank 3 Connect Tubing for Flushing Precharge Storage Tank 5 7 psi 0 34 bar 0 034 MPa 3 8 Tubing Product Water Out 1 3 2 4 3 8 Tubing To Storage Tank 3 8 Tubing Supply Water In CLOSE OPEN 4 Prepare to Flush the Everpure 4FC5 Cartridge 1 3 2 Connect Electrical Install Bypass Plug CLOSE Unplugged Inlet Water ...

Страница 6: ...e 4FC5 Cartridge 1 2 4FC5 OPEN Plugged 1 4 Inlet Water Install Cartridge Flush to Drain Five 5 Minutes 2 3 5 Plug In Power Cord Unplug Power Cord Closed Valve As Shown 6 Prepare the Pentair GRO 75EN Cartridge for Use Replace Membrane 2 Remove Cover 1 ...

Страница 7: ...rtridge Storage Tank Not Shown 1 Remove Plug Push White 3 8 Tubing into White Connector 3 2 4 Push Red 3 8 Tubing into Red Connector Push Blue 3 8 Tubing into Blue Connector 8 Flush Pentair GRO 75EN Cartridge Plug In Power Cord Inlet Water Open Flush To Drain for 24 Hours 1 4 2 3 Fully CLOSE Turn To Right Unplug Power Cord ...

Страница 8: ...d Open 2 Solid Blue Blending Valve 1 4 Turn or Less 2 Solid Blue Blending Valve 1 4 Turn or Less 5 2 4 Repeat Until Flashing Blue Blending Valve 1 8 Turn Press Press 1 3 6 Plug In Power Cord Unplug In Power Cord OPEN One 1 Full CCW Turn of the Valve ...

Страница 9: ...9 EVERPURE CONSERV LT S HIGH EFFICIENCY RO SYSTEM INSTALLATION AND OPERATION GUIDE 10 Install Everpure 4CC Cartridge 1 2 4CC 1 Remove Port Plug 2 Install 4CC Cartridge ...

Страница 10: ...10 WATER YOU CAN TRUST 11 Flush Storage Tank and 4CC Cartridge 3 2 CLOSE OPEN 1 When Pump Turns Off OPEN valve 5 4 Plug In Power Cord Repeat Steps 2 and 4 Four 4 Times ...

Страница 11: ...more water the system is required to produce the longer the membrane will last INSTALLATION NOTE Please read all instructions specifications and precautions before installing and using your Conserv LT S Processor NOTE The Conserv LT S Processor may be installed hanging on a wall or set under a sink NOTE The Conserv LT S Processor must be installed vertically WARNING Improper connection of the appl...

Страница 12: ...er end to the Storage Tank and insert it into the ball valve The ball valve should be closed 5 Connect a length of 3 8 O D tubing to the Everpure Conserv LT S Processor outlet connection Secure an open ball valve to the other end and position it over a drain Water Supply Inlet Figure 3 4 Prepare to Flush the Everpure 4FC5 Cartridge The Everpure 4FC5 Filter Cartridge is installed and flushed before...

Страница 13: ... connect ports that are labeled Inlet White Product RO Water Blue Drain Red White Port Figure 7 CAUTION DO NOT remove the quick connect elbow fitting from the red tubing This assembly contains a flow control that may become damaged if removed The efficiency of the Everpure Conserv LT S Processor may be affected 1 Remove the GRO 75EN RO Membrane Cartridge from its packaging and position it with the...

Страница 14: ... depress blue cap Do not submerge entire unit 4 One 1 or two 2 lights will light immediately Match light s to the chart to get approximate range of Total Dissolved Solids TDS Release the blue cap and write down the TDS range 0 75 ppm solid blue 76 150 ppm flashing blue 100 250 ppm flashing blue solid green 251 300 ppm solid green 301 400 ppm flashing green 401 500 ppm solid green flashing green 50...

Страница 15: ...ck the tank air pressure Ensure that the tank has 5 7 psi 0 34 bar 0 034 MPa of precharge upon installation Add air if needed Tank pressure may be checked using an air pressure gauge on the air valve on the side of the tank Blended Water Outlet Figure 12 1 OPEN the ball valve on the top of the Storage Tank 2 CLOSE the ball valve at the blended water out tubing 3 Apply power It will take approximat...

Страница 16: ... access to the flow controller and the attached capillary tube Replace the flow controller Figure 13 in the red tubing 6 Insert a new flow controller Figure 13 with capillary tube into the tubing Replace the quick connect elbow Flow Controller Red Tubing White Tubing Blue Tubing Figure 13 7 Insert the new GRO 75EN RO Membrane Cartridge into position in the clip Rotate the cartridge so the quick co...

Страница 17: ...high quality silicone lubricant Replace the central plug assembly and insert the flushing cartridge into the left side head for the Everpure 4FC5 Cartridge NOTE Do not use petroleum jelly to lubricate the o rings It will degrade the o rings 2JT Plug 4CC Port Connect White and Blue Tubing Cartridge with Sanitizer Figure 15 6 Install a plug into the port for the 4CC Filter Cartridge NOTE A plug was ...

Страница 18: ... 3 Unscrew the tank valve from the storage tank 4 Inspect the rubber seal inside the threaded area Repair or replace the valve as needed 5 Thread the tank valve onto the top of the tank opening by turning it clockwise until snug 6 Wet the tubing and insert it 3 4 1 9 cm into the quick connect fitting 7 Turn water supply ON and CLOSE the application valve 8 Allow processor to pressurize for several...

Страница 19: ...nstallation must comply with state and local regulations Systems certified for cyst reduction may be used on disinfected waters that may contain filterable cysts Substances that may be reduced are not necessarily in your water Filter must be maintained according to manufacturer s instructions including replacement of filter cartridges The efficiency rating for this system is 56 31 Efficiency ratin...

Страница 20: ...re technical service at the phone number listed below Any part found materially defective within the terms of this warranty will be repaired or replaced at PFS s discretion by your local dealer or Pentair Everpure technical service You pay only freight from our factory and local dealer charges Any item repaired or replaced pursuant to this warranty will be covered under the original warranty terms...

Отзывы: