9
THAW COMPLETE MODE
When the
Thaw
feature is used, the feeding should be removed immediately
to prevent over-warming.
The warmer emits a ready tone.
The heater immediately shuts off, and milk must be refrigerated.
WARNING: Milk must be refrigerated immediately.
The screen background will turn green and the following phrases will appear on
the screen:
OVERTIME MODE
After 30 minutes in
Ready to Feed
, the warmer will enter
Overtime
.
Warming and vibration will turn off.
The screen background will turn red and the word "Overtime" will appear in the
middle of the screen:
The Overtime screen will remain on until the unit is turned off or is returned to
Standby
by pushing the On/Off button or Enter button.
Note: As the manufacturer, we recommend not using the feeding after
the 30 minute hold time, but that feeding should be handled according
to the hospital policy.
DISPLAYED ERRORS
If the Penguin Nutritional Warmer encounters an error, the error will be displayed
on a red screen with a warning tone. If this should occur, discontinue use of the
warmer and call Ameda CustomerCare at 1.877.99.AMEDA (26332).
1. At the beginning of a shift, take out one Therma-Liner per user.
2. Using a permanent marker, enter the following information onto the Therma-Liner:
a. Patient Information b. Date/Time c. Information
3. Open the water compartment (blue zipper) and fill to water line.
4. Pinch one end of the blue zipper between your thumb and forefinger and slide
down the length of the zipper to close the water compartment.
5. Repeat to ensure a proper seal. This compartment will remain sealed for the
duration of the 12 hour shift.
Note: Use care when removing and re-inserting milk container to
prevent damage to interior of water compartment.
4.4 THERMA-LINER PREPARATION