11
6.2 THAW FROZEN MILK TO COOL LIQUID STATE
CAUTION: THE PENGUIN NUTRITIONAL WARMER IS NOT RECOMMENDED FOR
THAWING LESS THAN 15 ml of BREAST MILK IN STORAGE BAGS!
1. Scroll to choose the container type and size of interest.
2. Press Enter.
3. Only when ready to begin warming the feeding, place the feeding container into
the inner compartment (orange zipper) of the Therma-Liner and push it to the
bottom. Press firmly along the entire orange zipper to encompass the container as
much as possible. Use care when removing and re-inserting feeding containers to
prevent damage to the interior water compartment.
4. Place the prepared Therma-Liner in the well with the sealed opening at the top.
5. Press Enter*.
6. The device will begin warming, vibrating, and alternating screens between the profile
selected and elapsed time.
7. When the feeding has been warmed to a liquid cool state, the device will enter
Thaw Complete mode
.
8. Please immediately remove the feeding** from the Therma-Liner and set the
Therma-Liner aside until next use during the shift.
9. Press On/Off to return the warmer to
Standby mode
.
*Note that if a wrong entry was made, you may press the Cancel selection button to
go back a screen.
**If there are still some frozen particles in the feeding, let it sit until the remainder
thaws, or rotate the container gently to thaw the rest of the way. It is better to have
a few remaining ice crystals than to overshoot the refrigerated range. Any portion
that is meant for immediate feeding may be placed in a bottle or syringe for further
warming. Feeding for the future should be placed immediately in the refrigerator for
up to 24 hours. After 24 hours, it should be discarded.
7. CLEANING
1. Disconnect Power Cord from the wall receptacle.
2. Prepare reagent wipes (e.g. CaviWipe™ or SaniWipe™) containing
Isopropanol, Ethylene Glycol Monobutyl Ether (2-Butoxyethanol),
Diisobutylphenoxyethoxyethyldimethyl benzylammonium chloride.
3. Using the prepared reagent wipe, clean and remove all surface debris from the
device, including the power cord.
4. Repeat as deemed necessary using a newly prepared reagent wipe each time.
5. After the last cleaning wipe is used, allow device to air dry for a minimum of 10
minutes or dry with a soft cloth before returning the device to service.