Operators Manual Model PT2000/H Series Gas Monitor Rev 2. January 2014
7.3 Power Up Wireless Detectors
If the Detectors are solar powered then simply connect the power cable from the solar
array module to the power plug receptacle on side of detector as shown in Figure 8 in
section 7.1.
However, for the Toxic detectors that are internal battery powered an internal Power
Switch must be activated to Power On the detector.
First Time Detector Power Up
The detectors from the factory are shipped with batteries installed but the ribbon
cable inside disconnected. If the detector is powered up for the first time then follow
the steps below before proceeding to the next step
Remove the Glass Cover.
Loosen two (2) thumb screws on top of the transmitter board assembly.
DO
NOT COMPLETELY REMOVE THE THUMB SCREWS.
Find a short ribbon cable attached on the connection board assembly next the
battery compartment.
Connect the transmitter board assembly to the connection board via this
ribbon cable.
Place the transmitter board back on the stand offs on the connection board
and secure the board with thumb screws. Do not Over Tighten the thumb
screws.
Power On / OFF Detectors:
Place the magnetic wand on top of the sensor enclosure to activate the internal
power switch. Remove the magnet when sensor startup messages appear on
the LCD.
To shut down the detectors when not in use place and
hold the magnet on top of the enclosure just like in
power On mode for few seconds. Detector will display
shut down message and will shut down the power from
the batteries in 5 seconds.