P/N UM500KL 500KG REV20191120_Web
14
SET UP INSTRUCTIONS
To enter the Options mode, push and hold the
ENTER
button while pressing and releasing the button
.
Auto Off Function
When powering the scale via the AC adapter, the display can be set to power off after a period
of inactivity.
The Auto Off default setting is “disabled”. Follow this procedure to enable or disable
the Auto Off function.
Note:
When powering the scale via batteries, the Auto Off function is
automatically enabled.
Enable / Disable Auto Off
1. While the scale is turned off, press and hold the
ENTER
button then press and release the
button. Hold the
ENTER
button until “AOF=d” or “AOF=E” appears on the display, then
release. Press the
ENTER
button again to change the setting.
AOF=d
Disables the Auto Off function
AOF=E
Enables the Auto Off function
2. Press
CLEAR/REWEIGH
to save your selection and exit the Options mode.
3. The scale will restart. The display will show
“UEr”, followed by a version number, and then
dashes. When the display shows “0.0” the scale is ready to use.
Setting Auto Off Time
When powering the scale via batteries or if Auto Off is Enabled for the AC adapter, the period of
inactivity before the scale powers off can be set to a specific amount of time. The default setting for
Auto Off time is 2 minutes. Follow this procedure to set the number of minutes of inactivity before
power off.
1. While the scale is turned off, press and hold the
ENTER
button then press and release the
button. Hold the
ENTER
button until
“AOF=d” or “AOF=E” appears on the display, then
release.
Use the ▲ or ▼ buttons to advance the display screen until “AFt” appears. Press
the
ENTER
button to change the Auto Off time.
AFt = 1
Scale will power off after 1 minute of inactivity
AFt = 2
Scale will power off after 2 minutes of inactivity
AFt = 3
Scale will power off after 3 minutes of inactivity
2. Choose the setting you prefer and press
CLEAR/REWEIGH
to save your selection and exit
the Options mode.
3. The scale will restart. The display will show
“UEr”, followed by a version number, and then
dashes. When the display shows “0.0” the scale is ready to use.
Setting Sound Options
The display can be set to stay silent or beep while operating the scale. The default setting for
Sound is “enabled”. Follow this procedure to enable or disable sound.
1. While the scale is turned off, press and hold the
ENTER
button then press and release the
button. Hold the
ENTER
button until
“AOF=d” or “AOF=E” appears on the display, then
release.
Use the ▲ or ▼ buttons to advance the display screen until “Snd” appears. Press
the
ENTER
button to change the setting.
Snd = E Enables sound so the scale beeps when keys are pressed.
Snd = d Disables sound so the scale stays silent when keys are pressed.
2. Choose the setting you prefer and press
CLEAR/REWEIGH
to save your selection and exit
the options mode.
3. The scale will restart. The display will show
“UEr”, followed by a version number, and then
dashes. When the display shows “0.0” the scale is ready to use.