P/N UM500KL 500KG REV20191120_Web
16
SET UP INSTRUCTIONS (CONTINUED)
pressing
ZERO / TARE
. Users will see dashes on the screen as the scale is
finding zero.
PU2 = E Enabling this feature means the scale will zero itself when it is powered on. After
each use, the scale will clear the weight of the patient and set itself back to the
initial zero level it found on start-up. Health o meter® Professional recommends
keeping this setting as Enable.
2. Choose the setting you prefer and press
CLEAR/REWEIGH
to save your selection and exit
the Options mode.
3. The scale will restart. The display will show
“UEr”, followed by a version number, and then
dashes. When the display shows “0.0” the scale is ready to use.
Everlock
®
(500KL only, this feature is not available on 500KG)
The Everlock
®
feature allows the measuring unit to be permanently locked on the scale, disabling
the
UNIT
button. The Everlock
®
default setting is “disabled”. Follow this procedure to engage
Everlock
®
.
Warning: This lock is permanent and cannot be reversed.
Notes:
Press
CLEAR/REWEIGH
at any time to exit the Everlock
®
procedure.
Once Everlock
®
is engaged the Unt and UdP options will no longer be displayed when
scrolling through the Options menu.
There is no method to undo Everlock
®
without sending the scale back to the manufacturer.
1. While the scale is turned off, press and hold the
ENTER
button then press and release the
button. Hold the
ENTER
button until
“AOF=d” or “AOF=E” appears on the display. Use
the ▲ or ▼ buttons to advance the display screen until “ELC” appears. Press the
ENTER
button to change the setting.
ELC =d
Default setting with Everlock
®
disabled.
ELC = E Engages Everlock
®
. Once this option is selected, the display will prompt you to
begin the locking procedure. If this is desired, follow the instructions below.
2. After Everlock
®
is selected,
“PASS1” will appear on the display. This prompts the user to
enter a password. Press the following keys in this sequence:
BMI
,
ENTER
, ▼, ▲
3.
If entered correctly, “ELC=L” (if scale was in LB mode) or “ELC=6” (if scale was in KG mode)
will appear on the display.
4.
Use the ▼ and ▲ keys to toggle between (L) Pounds and (6) Kilograms. When the desired
unit is displayed, press
ENTER
and “PASS2” will appear on the display.
5. This will prompt the user to enter the second password. Press the following keys in this
sequence:
▲, ▼,
ENTER
,
BMI
6. If
entered correctly, the scale will display “ELC = L” or “ELC = 6,” depending on the unit of
measurement that was selected. Everlock
®