MU-Thermocouple1 CAN FD – User Manual
30
CAN (FD)
Supported bit timing values
Prescaler (BRP)
Time Segment 1 (TSEG1)
Time Segment 2 (TSEG2)
Synch. Jump Width (SJW)
Nominal
1 … 512
1 … 256
1 … 128
1 … 128
Data
1 … 32
1 … 32
1 … 16
1 … 16
Galvanic isolation
Pins CAN_H, CAN_L, and CAN_GND are decoupled
galvanically against the supply with 500 V
Internal termination
Switchable, not activated on delivery
CAN ID reserved for
configuration transfer
7E7h
Thermocouple Inputs
Count 8
Connectors
Mini sockets for thermocouples (DIN EN 50212)
Supported thermocouple
types (IEC 60584-1) with
measuring range
J = -210 … +1121 °C (-346 … +2050 °F)
K = -200 … +1370 °C (-328 … +2498 °F)
T = -200 … +400 °C (-328 … +752 °F)
Sampling rate
3 Hz
Measurement accuracy of
the measurement unit
0.2 % or 1 K
Influence of ambient
temperature
10.5 ppm/K
Reference temperature
sensors
4 (1 per measuring module)
Measuring accuracy of the
reference temperature
sensors
Typically: 0.5
K
Maximum values: ±1 K at 0 … 70 °C (32 … 158 °F)
±2 K at -20 … 85 °C (-4 … 185 °F)
±3 K at -40 … 125 °C (-40 … 257 °F)
Max. resolution for CAN
data
1/16 °C
Measuring category
CAT I (only electric circuits that are not connected to
the mains)
2
The values refer only to the accuracy of the measuring electronics and do not
consider the temperature sensors.