MU-Thermocouple1 CAN FD – User Manual
11
2.4
Status LEDs
The device has status LEDs for each of the eight thermocouple
inputs as well as for the power supply.
LED
LED Indication
Meaning
Lights red
An intact thermocouple is connected.
If the corresponding LED does not light up,
despite the thermocouple being connected,
the cable connection or the thermocouple
may be broken.
Thermocouple
inputs
(configurable)
Flashes red quickly
(4 Hz)
No thermocouple is plugged in, the
thermocouple is not connected correctly,
or the cable at the thermocouple is broken.
Flashes green
(1 Hz)
Normal operation
Flashes green
quickly (2 Hz)
Missing configuration.
Transfer a configuration to the measuring
unit via CAN.
Power supply
Flashes yellow
Firmware update in progress.
The listed LED indication of the thermocouple inputs corresponds to
the pre-configuration on delivery. The software Thermocouple FD
Configuration can be used to reconfigure the status LEDs of the
thermocouple inputs. More about this can be found in chapter 5 on
page 19.