MU-Thermocouple1 CAN FD – User Manual
22
6
Hardware Modifications
For special applications, various modifications can be done on the
circuit board of the MU-Thermocouple1 CAN FD:
Changing the device ID to enable the configuration of several
measuring units connected to the same CAN bus: chapter 6.1
below.
Activation of the internal termination: chapter 6.2 on page 24.
6.1
Device ID
If you want to configure several measuring units on the same CAN
bus, you must assign each measuring unit its own ID in the range 0
to 15. This is done by means of a rotary switch on the board. This
allows the measuring units to be distinguished by the configuration
software.
Do the following to set the ID of the measuring unit:
Attention!
Electrostatic discharge (ESD) can damage or destroy
components on the card. Take precautions to avoid ESD.
1.
Disconnect the power supply.
2.
Remove all connected cables from the measuring unit.
3.
Remove the two screws of the D-Sub connector and the four
screws on the front panel to remove it.
4.
Remove the two upper screws of the back panel and remove
the housing cover.