MU-Thermocouple1 CAN FD – User Manual
12
3
Operation
3.1
Safety Instructions
Attention!
Danger due to electric shock! Risk of destroying the
measuring unit!
You may only measure temperatures on energized parts when
these are not directly connected with the mains voltage
(measuring category CAT I). The measuring unit must not be
used in the measuring categories CAT II, CAT III, or CAT IV.
Never apply a voltage higher than 30 V between thermocouples
or between a thermocouple and earth.
Attention!
Risk of burns!
At ambient temperatures of 70 °C (158 °F) and above a
protection against contact must be ensured for the measuring
unit, i.e. the surface may no longer be tangible.
3.2
Operation with Default Configuration
The measuring unit is delivered with a standard configuration which
allows you to immediately measure and record the measurement
data via CAN without further adjustments. The CAN communication
is set to the CAN 2.0 protocol and to a bit rate of 500 kbit/s. More
information about creating your own configuration can be found in
chapter 5 on page 19.