Installation
Generator Environment
The generator should, if possible, be installed next to the application,. If, however, this is
not possible, please consider the distance between the generator and your application,
since pressure drops can result from extended lengths of tubing and may affect gas
delivery.
Performance of the generator is affected by ambient conditions. Note should also
be taken to the proximity of Air Conditioning outlets. Air conditioning systems can
sometimes give rise to “pockets” of air with high relative humidity. Operation of the unit
within such a pocket could adversely affect its performance. Consideration should also
be given to the air flow around the unit. It is recommended that an air gap of 75mm (3”)
should be maintained down both sides and at the rear of the unit. Please refer to the
drawing below for the general dimensions of the unit.
Please ensure Generator is situated in a well ventilated environment and is positioned to
allow easy disconnection from the mains supply if required.
Minimum Operating Ambient Temperature:
5 °C (41 °F)
Maximum Operating Ambient Temperature:
35 °C (95 °F)