Page 10
Technical Specification
TOC 1000
Environment
Minimum Operating Ambient Temperature
5°C (41°F)
Maximum Operating Ambient Temperature
35°C (95°F)
Maximum Altitude
2000m
Maximum Relative Humidity
70% @ 35°C
Minimum Storage Temperature
*
5°C ( 41°F)
Maximum Storage Temperature
*
35°C (95°F)
Maximum Storage Temperature
*
90% non-condensing
Inlet Conditions
Min/Max Inlet Air Pressure
100 psi/6.9 bar - 145 psi/10.0 bar
Minimum Air Inlet Flow
6 L/min
Minimum Air Inlet Quality
ISO 8573 - 1:2010 Class 1.4.1
†
Maximum Inlet CH
4
Concentration
100ppm CH
4
Generator Outlets
Maximum Gas Output Pressure
50 psi / 3.4 bar
Maximum Outlet Flow Rate
**
1000 cc/min
Gas Purity
<1ppm CO2 <1ppm CO <1ppm SOx
<1ppm NOx<0.1ppm CH4
Dewpoint (°C)
-60°C / -76°F
Particles
<0.01 μm
Phthalates
Phthalate & BHT Free
Suspended Liquids
None
Gas Outlets
1 x 1/4” BSPP x 6mm push-fit connection
Start-Up Time
120 minutes
Electrical Requirements
Voltage
120V±5% / 230V ±10%
Frequency
120V/ 60Hz / 230V 50/60Hz
Current
120V 4A / 230V 1A
Power Cord (Supplied)
C13 Plug
Circuit Breakers
10A
Pollution Degree
2
Insulation Category
Class 1
Transient Over Voltages
Over Voltage Category II
General
Dimensions mm (inches) HxWxD
343 x 270 x 530 mm (13.5 x 10.7 x 20.9'')
Shipping Dimensions mm (inches) WxDxH
435 x 650 x 400mm (17.1 x 25.5 x 15.7")
Generator Weight Kg (lbs)
27kg (59.6lbs)
Shipping Weight Kg (lbs)
38kg (83.8lbs)
Noise Level
Virtually Silent
* Note: Please ensure generator is situated in a well ventilated environment.
** Note: Flows in LPM are expressed as normalised volumes at 101.3kPa, 20°C.
†
This unit should be connected to an air supply that, as a minimum, meets ISO 8573-1:2010 Class 1.4.1 or connected to the relevant Peak air compressor model.
Содержание TOC 1000
Страница 25: ...Page 25...