Page 17
Air Purity
The Precision Zero Air Series generator should be connected to an air supply that, as a
minimum, meets ISO8753-1:2010 Class 1.4.1
Class 1 Particulate
In each cubic metre of compressed air, the particulate count should not exceed 20,000
particles in the 0.1 - 0.5 micron size range, 400 particles in the 0.5 - 1 micron range and
10 particles in the 1 - 5 micron size range.
Class 4 Water
A pressure dew point (PDP) of +3°C or better is required and no liquid water is allowed.
Class 1 Oil
In each cubic metre of compressed air, not more than 0.01mg of oil is allowed. This is a
total level for liquid oil, oil aerosol and oil vapour.