background image

Precision Hydrogen SL 

(All Models)

User Manual

Copyright © Peak Scientific Instruments Ltd 2020 — Precision Hydrogen SL User Manual Rev 1 RSID 3516 EN —10/04/2020

Содержание SL 100 cc

Страница 1: ...Precision Hydrogen SL All Models User Manual Copyright Peak Scientific Instruments Ltd 2020 Precision Hydrogen SL User Manual Rev 1 RSID 3516 EN 10 04 2020...

Страница 2: ...tings Kit Contents 21 Unpacking Instructions 21 Installation 22 Location 22 Generator Overview 23 General Dimensions 23 Unit Controls 24 Rear Connections 24 Tubing 25 H2 Ventilation 25 Water Quality 2...

Страница 3: ...hard copy version can refer to the contents page to find the relevant information Users of the soft copy version can use the hyperlinks from the contents page as well as the hyperlinks between sectio...

Страница 4: ...on Hydrogen SL is based on proven technology producing hydrogen by electrolysis using a Proton Exchange Membrane PEM cell Moisture removal is achieved by a two stage patented Nafion technology and des...

Страница 5: ...on Temperature Over 220 C Vapour Pressure 25 C 17mm CFH 20 C Water Viscosity 25 C N A Solubility in Water Negligible Specific Gravity 25 C 1 1 1 3 pH 25 C 7 9 5 in aqueous slurry Eye Protection Safety...

Страница 6: ...tions If irritation persists move patient to fresh air Ingestion If resin has been confined to the mouth administer 250 ml of water as a mouthwash Avoid swallowing This product is considered non hazar...

Страница 7: ...Odourless Explosion Limits N A Decomposition Temperature Over 155 C Vapour Pressure 25 C 17mm CFH 20 C Water Viscosity 25 C N A Solubility in Water less 1 0 in weight pH 25 C 2 10 at 5 w w in water E...

Страница 8: ...Remove from source of exposure Ingestion Wash out mouth with water If large amount swallowed or symptoms develop obtain medical attention Toxicological Data The lethal dose for humans for synthetic a...

Страница 9: ...Odourless Explosion Limits N A Decomposition Temperature N A Vapour Pressure 25 C N A Viscosity 25 C N A Solubility in Water Insoluble pH 25 C N A Eye Protection Safety Glasses BS2092C Storage All co...

Страница 10: ...Remove from source of exposure Ingestion Wash out mouth with water If large amount swallowed or symptoms develop obtain medical attention Toxicological Data The lethal dose for humans for synthetic a...

Страница 11: ...in terms of clause 2 hereof the Customer must 1 Reasonably satisfy the Company that the Goods have been properly installed commissioned stored serviced and used and without prejudice to the generalit...

Страница 12: ...ions are fully understood or met A CAUTION notice denotes a hazard It calls attention to an operating procedure process or similar which if not correctly performed or adhered to could cause damage to...

Страница 13: ...ty void 3 If you are using a manual fill system the tank filter must be replaced after filling the tank with water 4 If your system requires an external tank do not place the external water bottle abo...

Страница 14: ...10 Electrical Equipment for measurement control and laboratory use Electromagnetic Compatibility Directive 2014 30 EU EN 61326 1 2013 Electrical Equipment for measurement control and laboratory use FC...

Страница 15: ...2 19 EU These obligations are being met within the B2B compliance group Peak Scientific Instruments Ltd has developed all reasonable due diligence controls to ensure that our products comply with the...

Страница 16: ...i 6 9 bar Electrical Requirements Voltage 100 240VAC 10 Frequency 50 60 Hz Current 6A Max Input Connection C14 Connection Power Cord Supplied C13 socket to local connection 10A minimum Pollution Degre...

Страница 17: ...round the external electrical circuit These protons and electrons then recombine at the cathode of the electrolyser again with the aid of a precious metal catalyst to produce the hydrogen gas The high...

Страница 18: ...ml at a pressure of 7 6bar 110psi The unit also meets the European Electromagnetic Compatibility and Low Voltage Directives The Generator is CE approved and has the CE mark attached All of the Precisi...

Страница 19: ...ion Hydrogen range have the following safety features to ensure safe and reliable operation Overpressure switch Safe and accurate control of pressure Internal leak detection Cell shut down in critical...

Страница 20: ...ier and Peak Scientific or the Peak Partner from where the unit was purchased Follow the unpacking instructions posted on the side of the crate Please save the product packaging for storage or future...

Страница 21: ...x 1 8 Installation Guide Precision Hydrogen x 1 9 4 Litre Water Bottle x 1 10 A5 simple user maintenance guide x 1 11 Desiccant Cartridges x 1 100cc x3 200cc 12 DI cartridge x1 to be installed prior...

Страница 22: ...ade to allow the volume of air in the room to be changed 5 times per hour The risk with hydrogen is a leak into the environment raising the hydrogen content to the explosive limit The rear of the gene...

Страница 23: ...23 WARNING Generator Overview General Dimensions The Generator must always be placed on a flat level surface Failure to do so will affect the performance of the Generator 160 mm 6 3 385 mm 15 2 250mm...

Страница 24: ...Page 24 Unit Controls Rear Connections Start Button Power Switch and Inlet H2 Output H2 Vent Water Fill autofill and pressurized fill units only Water Drain...

Страница 25: ...ommends that the H2 Exhaust port at the rear of the unit should be connected to an exhaust hood or other ventilation system to ensure safe disposal of vented gasses Tubing sizes for the vent port are...

Страница 26: ...with DI water to be functional For manual fill systems follow the below steps 1 Lift the hinged lid back then remove the tank cover 2 Fill the water tank from your lab s supply of DI water and ensure...

Страница 27: ...rom the fittings kit attach the barb to the tubing Then connect one end to the water supply and the other to the Fill port at the rear of the Precision Hydrogen For efficient water filling the externa...

Страница 28: ...ce the low level water sensor disengages the generator will then begin producing hydrogen as before Four and eight litre water storage bottles can be purchased from Peak Scientific please contact your...

Страница 29: ...rate of the Precision Hydrogen Generator is completely dependent on the rate at which the hydrogen is produced The graphs below show an estimation of the water consumption from the generator in Stand...

Страница 30: ...ified electrician If a substitute main supply cord is used ensure that it has adequate ratings and appropriate approvals for the country of operation Failure to do so could cause damage to the generat...

Страница 31: ...ill cause the generator to cease hydrogen production illuminate the water tank blue and exhaust any stored hydrogen through the exhaust port at the rear of the generator Note If leak test is unsuccess...

Страница 32: ...ed off by the power switch on the rear panel Cleaning Clean the outside of the generator using only warm soapy water and a clean damp cloth Ensure the cloth is thoroughly rung out to remove excess flu...

Страница 33: ...e is put in by lowering into the recess and twisting 90 clockwise to lock in position 1 Depressurise the generator and disconnect power To replace the DI cartridge lift the hinged lid back then remove...

Страница 34: ...side of the generator The capsule should be replaced when all of the silica gel has visibly darkened 1 To replace the capsule depressurise the generator and disconnect power Lift the hinged lid back t...

Страница 35: ...Page 35 3 The cap can then be replaced ensuring not to over tighten the cap and the hinged lid can be closed 4 The unit can now be switched on...

Страница 36: ...lans Peak Scientific offer two service plans The Complete Service Plan specifically designed for generators operated in critical environments also includes full breakdown cover guaranteed response tim...

Страница 37: ...ur service provider The generator is illuminated solid ORANGE Contact your service provider The generator illumination is pulsing orange Ensure the water tank is adequately filled Check for external l...

Страница 38: ...Scientific product from the date of installation Once registered the warranty will be honoured for a period of 12 months If you wish to defer the installation of your generator you must notify Peak Sc...

Страница 39: ...Page 39...

Страница 40: ...y leading rapid response service to our customers With Peak Protected your laboratory s productivity becomes our top priority To discuss Peak Protected generator cover and payment options speak to you...

Отзывы: