Infinity 9000
User Manual
Page 13
2014 © Peak Scientific - Rev. 2 - 28/05/14
RAC Filter Element Replacement
Ensure that Generator is in Safely Isolated Condition
1.
Open the filter bowl valve to ensure it
is depressurised.
2.
Remove the filter bowl by turning it
as shown to reveal the RAC filter
element Part number
00-0041.
a.
For easier access, the filter
assembly can be removed
from the cabinet
by removing the 4 of nyloc
nuts and disconnecting the
tubing.
3.
Remove and replace the filter
element and re-assemble.
00-0041