Air Connection
Infinity 9000
Generator should be connected to a clean, dry, OIL - FREE source
of compressed air. A minimum inlet pressure of 100 psig (6.7 barg) is required.
Any doubts as to the suitability of your compressed air supply should be
referred to Peak Scientific or any of their authorised partners.
The generator has a Breathing Air Filter with ¼” BSPT connection to the left
side of the unit. The Compressed Air supply should be connected here. This
filter will drain moisture and is equipped with an automatic drain. The drain
should be led to a convenient place. Nitrogen outlet is via an Activated Carbon
reverse-acting filter mounted on the right side, again with a ¼” BSPT
connection for outlet.
Fittings Kit
Supplied in the Fittings Kit are all the fittings required to connect the Infinity
9000 Generator to the application. The contents of the Fittings Kit are as
follows:
1.
6mm Push Fit Tube Fitting
× 2
2.
6mm Tubing
× 3m
3.
N8 Supa Counter Sunk Screw
x 2
4.
Raw plug for No.8 Screw
x 2
5.
User Manual – Infinity 9000
× 1
All of the Generator output ports are located on the output panel at the rear of
the Generator.