Page 21
Service Requirements
Service Schedule
Purchase Interval Component
Qty.
Visit
12 Months
Genius SQ 24 Annual Maintenance Kit
1
www.peakscientific.com/ordering
Compressor
Assembly *
Compressor Assembly Genius SQ 24
2
* If your compressors have been refitted 3 times they must now be replaced. Please
note that when replacing the compressor assembly, the full Annual Maintenance Kit will
not be required. Please contact your service provider for more information.
Note: Only Peak approved components should be used.