Page 16
WARNING
Electrical Connection
Connect the generator to an appropriate 100, 120 or 230 volt single-phase supply, refer
to the generator serial plate for input specification and ensure your supply matches the
requirements.
If the appropriate power cord is not supplied; a new plug, rated to at least 12 amps, can
be fitted by a qualified electrician.
This unit is classified as SAFETY CLASS 1. THIS UNIT MUST BE
EARTHED. Before connecting the unit to the mains supply, please
check the information on the serial plate. The mains supply must be of
the stated AC voltage and frequency.
EARTH/GROUND (E):-
Green & Yellow or
Green
LIVE (L):-
Brown
or
Black
Neutral (N):-
Blue
or
White
Electrical requirements are 100, 120 or 230VAC n/- 10% depending on chosen
model. However, running continuously at voltages outwith this is not recommended.
Extended periods at extremes can have a detrimental effect on the operation and life of
the generator.
If the equipment is used in a manner not specified by the manufacturer,
the protection provided by the equipment maybe impaired.
If the
supply voltage
is 219V or less, a
dual tap transformer
must be fitted. This can be
directly ordered from Peak Scientific (
P/N 06-3200
).
If a substitute main supply cord is used, ensure that it has adequate ratings and
appropriate approvals for the country of operation. Failure to do so could cause
damage to the generator or risk of overloading of the power cord.
WARNING