Page 25

High Duty Indication

The generator has a ‘HIGH DUTY’ indicator on the front panel. This indicator monitors 

the running condition of the internal compressors and illuminates when they have been 

operating continually for a period of 8 hours.
There are a number of extreme conditions that can affect the duty cycling of the 

compressors inside the generator. These extremes include very high ambient 

temperatures, low supply voltages, a very high flow requirement and locations of high 

altitude. Any one of these or a combination of a number of these extremes can be 

evident in a customer site.
While the generator is designed to operate fully in these conditions it should be noted 

that the effect of these extremes can force the compressors to run continually. The 

compressors are fully capable of running continually with no detrimental effect. 

High Duty Indication Reset

If the compressors have been operating continually for a period of 8 hours the indicator 

light will be illuminated. Once the compressors return to a cycling mode the indicator 

light will automatically switch off. There is no manual intervention required.

Indication of Fault

In most installations the ‘HIGH DUTY’ indicator will not be illuminated. If your generator 

has been operating for a significant period of time, then suddenly develops a ‘HIGH 

DUTY’ indication with no apparent change in environmental conditions or flow 

requirements, it may be indicative of a problem with the system such as an external 

leak or imminent compressor failure.
If this is the case you should contact Peak Scientific or your service provider.
It should be noted that the generator will continue to operate so long as it can maintain 

pressure during this indication.

Содержание Genius SQ 24

Страница 1: ...Genius SQ 24 User Manual Copyright Peak Scientific Instruments Ltd 2019 GeniusSQ User Manual Rev 6 RSID 2630 EN 09 12 2019...

Страница 2: ...Overview 13 General Dimensions 13 Rear Connections 14 Unit Controls 14 Drain Connection 15 Electrical Connection 16 Start Up Sequence 17 Connecting to the application 18 Tubing Lengths 18 Normal Oper...

Страница 3: ...s manual is intended for end users and has been written as a reference document where you can skip to the relevant information Users can refer to the contents page to find the relevant information Ple...

Страница 4: ...upply to the generator must comply with ISO 8573 1 2010 Class 1 2 1 at all times d your air compressor dryer filtration and oil removal systems must be deemed suitable for use by Peak or a Peak Partne...

Страница 5: ...nd conditions outlined in this warranty state ment in any respect or to create any additional obligations or liabilities for any party involved 6 This warranty statement supersedes any and all prior w...

Страница 6: ...ted conditions are fully understood or met A CAUTION notice denotes a hazard It calls attention to an operating procedure process or similar which if not correctly performed or adhered to could cause...

Страница 7: ...EN 61010 1 2010 Electrical Equipment for measurement control and laboratory use Electromagnetic Compatibility Directive 2014 30 EU EN 61326 1 2013 Electrical Equipment for measurement control and lab...

Страница 8: ...confirm that as a manufacturer of articles we conform to the European Parliament and Council Directives 2011 65 EU for the original 6 restricted substances listed in the Restriction of Hazardous Subst...

Страница 9: ...nvironment Generator Outlets Maximum Gas Output Pressure 8 bar 116 psi Maximum Outlet Flow Rate 24 L min 0 85 cfm Particles 0 01 m Phthalates NONE Suspended Liquids NONE Gas Outlets 1 x 1 4 BSPP Drain...

Страница 10: ...e Peak Partner from where the unit was purchased Follow the unpacking instructions posted on the side of the crate It will require two people to remove the unit from the shipping crate and to manoeuvr...

Страница 11: ...Teflon Tubing x 3m 3 6mm PE Tubing x 3m 4 1 4 Compression Fitting x 1 5 6mm Push Fit Fitting x 2 6 Flow Control Silencer x 1 7 8mm Hex Key x 1 8 UK Mains Power Cable x 1 9 EU Mains Power Cable x 1 10...

Страница 12: ...ning outlets These can sometimes give rise to pockets of air with high relative humidity Operation of the unit within such a pocket could adversely affect its performance Consideration should also be...

Страница 13: ...Page 13 Generator Overview General Dimensions The generator must always be placed on a flat level surface Failure to do so will affect the performance of the generator WARNING 600 mm 750 mm 610 mm...

Страница 14: ...rces and applications All Connections should only be carried out by trained personnel Generator must be switched off and unplugged prior to any cleaning or maintenance operations 1 4 Compression Fitti...

Страница 15: ...connection or container It should be noted that the Generator can expel a considerable amount of water from this dependant on ambient humidity If a container is used it should be emptied at regular i...

Страница 16: ...Green LIVE L Brown or Black Neutral N Blue or White Electrical requirements are 100 120 or 230VAC nominal 10 depending on chosen model However running continuously at voltages outwith this is not reco...

Страница 17: ...d in the specifications Once this pressure is reached the compressors will continue to run until the internal tank upper pressure limit is reached and has been maintained for a period of time The syst...

Страница 18: ...tween the generator and the application can lead to a reduction in efficiency Tubing Lengths The diameter of the tubing which will be connected to the gas outlet is important and is determined by the...

Страница 19: ...tomatically produce the factory set flow and pressure as detailed in the Technical Specifications On Demand Gas The generator produces gas on demand If the application is operating and requires a gas...

Страница 20: ...compressors will decrease the rest period in the compressor cycle will be longer If the generator is installed in an extreme environment or is subjected to low supply voltage or high altitude the com...

Страница 21: ...ordering Compressor Assembly Compressor Assembly Genius SQ 24 2 If your compressors have been refitted 3 times they must now be replaced Please note that when replacing the compressor assembly the fu...

Страница 22: ...nce The generator will continue to operate as normal with the LED on Stage 2 If the service is not completed the generator will continue to run After 2 weeks the service LED indicator yellow will star...

Страница 23: ...to install and setup your generator IQ OQ Certified assurance for applications requiring documented qualification Complete plan Swift response by a Peak Service Engineer within 72 hours planned preve...

Страница 24: ...he cloth prior to use Cleaning should only be undertaken with the power switched off and the power cord removed from the rear of the generator Under no circumstances should any solvents or abrasive cl...

Страница 25: ...lly The compressors are fully capable of running continually with no detrimental effect High Duty Indication Reset If the compressors have been operating continually for a period of 8 hours the indica...

Страница 26: ...ider Yellow SERVICE LED on front panel is on constantly A compressor s due for service Contact your service provider Refer to Service Indication section of this manual for further information Yellow S...

Страница 27: ...ic com For generators that remain unregistered after 1 month from the shipment date the warranty will be considered active from the date of factory dispatch Complete Plan only Go Online or Complete an...

Страница 28: ...iority To discuss Peak Protected generator cover and payment options speak to your local Peak Representative or for further information contact protected peakscientific com Peak Scientific Fountain Cr...

Отзывы: