© PCE Instruments
26
Step Time (s)
Enter the time for a single step in seconds
Range:
1 … 9999
This parameter is enabled only for Auto Step Mode
(Manual CheckBox is Un-Checked)
Step Definition
Step Definition for the Step function
Available Options:
Temperature (Appear only if TC Display mode is Actual)
Percentage (Appear only if TC Display mode is
Percentage) User Defined
Step
Step Value in Temperature/mV/% according to TC
Display Mode and TC unit
Only appears if Step Definition is Temperature or
Percentage.
Define Steps
User Defined Step value for Manual and Auto Step
Mode.
This option appears only if Step Definition is User
Defined. Maximum 10 step value can be configured.
First enter the no. of step and then define step value in
serial order.
Repeat Format
How the stepping should be done.
Available Options:
UP
DOWN
UP/DOWN
DOWN/UP
This parameter is enabled only for Auto Step Mode
(Manual CheckBox is Un-Checked)
Repeat
Repeat Counts
Defines how many times the steps are repeated
Range:
1 … 9999
This parameter is enabled only for Auto Step Mode
(Manual CheckBox is Un-Checked)