© PCE Instruments
10
1
Time in HH:MM:SS format
Available in two formats
1. 24 hour (default)
2. 12 hour
This setting is available in Date/Time in Settings Men
2
Error Code Indicator
This icon is visible if any on-board peripherals like RTC, ADC, DAC, etc. are not working
properly, see chapter 5 Maintenance and Troubleshooting.
The List of error codes available in this device is given below.
Error code
Description
0
Memory corrupted or device unable to read/write it
1
RTC not working properly
2
Device unable to read battery information
3
Measure mode not working
4
Device unable to get source feedback reading
5
Data log memory corrupt
6
Source mode not working
9
More than one error from above list is occurring
3
USB Connection Status Icon
Icon is visible if USB charging adaptor or USB data cable is connected to the device. Icon
is different for both indications, see below.
USB data cable is connected and communication with PC
is available
USB charging adaptor is connected.
Battery starts charging.
4
Battery Charge Percentage Indicator
Always visible in Run page; battery % is shown in the centre of the icon and the icon
background is filled with green, yellow or red color if battery % is >= 50%, >= 20 and <20
respectively.
5
Data Logging Enable Status Indicator
Icon is visible if data logging is enabled and will flash when a data log is stored to the
memory