12 Total reset to factory settings
With this function, you can reset the gauge to factory settings.
That means:
a) all saved readings, statistics, calibration values, limit values and zero offset values will be
deleted (both F and N mode)
b) works calibration is activated (both F and N mode)
c) continuous mode is deactivated (both F and N mode)
d) the gauge is set to automatic FN identification (if available)
e) all options are reset to the first menu item, e. g. “μm/mm” or “Beeper ON”
Procedure:
1. Use On/Off key to switch off the gauge; a probe must be connected
2. Press and hold ESC key
3. Use On/Off key to switch on the gauge; you will hear a longer lasting beeptone; the welcome
screen comes up, followed by the start screen with the four dashes – – – –