17
down, users could press
<
Setting >
key to enter the
setting interface.
b.
On the setting interface users could select
<
Settings >
to
enter the system setting page, where the
< Fast boot >
is shown,
select the
< Fast boot >
, the calibration time will be 15s;
Unselect the
< Fast boot >
, the calibration time will be 60s.
c.
Press the
<
Setting >
key again to return to the previous
interface and the calibration time counting down restarts. The
detector will start the calibration, after this it will enter the test
interface.
Reminding
:
a.
The calibration pre-conditions of the detector are: there is no
ATP Quickswab within its test chamber and its top cover is
properly closed.
b.
If the
icon and
< Calibration is halted! >
are displayed on
the screen of detector, as shown in figure 2.5-1. Please take out
the ATP Quickswab from the detector to continue the
calibration process.
c.
If the
icon and
< Calibration is halted! >
are displayed on
the screen of detector, as shown in figure 2.5-2. Please close the
top cover again to continue the calibration process.