NMS044 Reference Manual
Configuring the Gateway for Larson Davis Instruments
A-18
Step 13.
Select the
LAN
menu,
Ethernet
sidetab.
TAKE NOTE
After this change you will not be able to connect to the gateway
with a wired Ethernet connection. If you need to restore the wired connection
without connecting to the gateway through the cellular connection, do a hard
reset on the gateway. This resets all items to the factory defaults. If you want to
use a cellular connection once again, you will need to repeat the process in this
section from step 1.
Figure A-22
LAN Settings
Step 14.
In the
Ethernet Port Configuration
section, change the
Port 1 State
to
Disable
, and click
Apply
.
Step 15.
Click the
Reboot
button in the top right. The gateway saves your settings and
reboots. It may take up to 3 minutes for the reboot.