Conducting a Final Equipment Check
28
Step 8.
After a few seconds, a message appears indicating the measured difference and
a prompt to save the results. Click
Yes
to save the calibration, or click
No
to
reject it.
Figure 3-12
Measured Calibration Difference
Step 9.
Carefully remove the calibrator from the microphone.
Step 10.
Select the Calibration History tab to view the acoustic calibration or the
calibration check summaries.
Step 11.
When the calibration process is complete, assemble windscreen and birdspike
back on the microphone.
3.7 Conducting a Final Equipment Check
We recommend that you complete these steps prior to leaving the system.
3.7.1 Checking System Status Indicators
Figure 3-13 explains the system status indicators.
Figure 3-13
NMS System Status LEDs
1.
2.
1.
Verify that the system battery is charged or
charging.
The system battery is fully charged when the solar
charge controller LED is solid green. A blinking green
indicates it is currently charging.
LEARN MORE
All system battery status indications are
Table A.5 Genasun Solar Charge Controller LED
.
2.
Check the gateway status.
The RV50X gateway has 4 LED status indicators on the
side of the unit. From left to right, indicators inform
you of the gateway’s
Network
,
Signal
,
Activity
, and
Power
.