PBI-Dansensor A/S
COPYRIGHT ©
PBI-220047-F 01/2008
29
MAP Check 9000 - User Manual
8. Error indications / Error correction
8. Error indications / Error correction
In MAP Check 900X-1 and MAP Check 9001-4 all error indications are presented as
Er.xx
,
xx
being a specific error number. They are also presented in MAP Mon for Windows.
In MAP Check 900X-2 errors are indicated by the LED on the front flashing quickly. The error
type can be seen in “MAP Mon for Windows”.
The error numbers are grouped in the following way.
Error
numbers
Description
1 - 9
User error. The value entered is not within the valid range, etc.
10 - 19
When MAP Check 9000 is switched on, all setup values are read from a
FLASH.
Data have been read correctly but contain errors (e.g. outside valid range).
The data in question are reset (lost) and stored in the FLASH. Always check
the data in question in the setup menu.
20 - 29
Hardware error. A reading/writing error has occurred while reading/writing
setup values to the FLASH.
30 - 31
Possible hardware error. Mains supply line failure signal error.
32 - 33
Errors in connection with setting up extra equipment (errors in
Pa.3
)
40
Error in calibration
50 - 99
Fatal hardware errors. (Except error No. 54).
Problem
Possible Cause
Possible solution
The display fails to show
O
2
values although the
pump is running.
MAP Check 9000 is not
warm. If for instance the
del.
parameter has been
set to 240 sec. by
mistake, 4 min. will pass
before an O
2
value is
shown
Wait until MAP Check 9000 has
warmed up. If the unit is cold, this
will take 20 min. (less after
relatively short interruptions)
The display is not lit
MAP Check 9000 is not
supplied with the correct
voltage
Check connections and switches
The fuse in MAP Check
9000 is defective
Replace the fuse (see
specifications)
The display flashes
FL.ER
Bent hose
Straighten the hose or replace it
Defective pump
Replace defective pump
Flow too low
Adjust flow
Measuring gas filter
blocked
Inspect / replace measuring gas
filter
Incorrect setting of Min.
and Max. flow limits
Check setting in the setup menu