Ammonium ISE Probe

Replacement Items

4

012-15718A

Ammonium, you should use a 1 mg/L low Standard 

Solution and a 10 mg/L high Standard Solution. 

NOTE: Temperature compensation is not typical of 

ISE measurements - for best performance, try to 

ensure that the standards are within 5 to 10°C of the 

sample.

Measurement:

If using a measuring devicewith direct concentration 

reading capability (see the software on-line help or the 

measuring device’s instruction manual for specific 

information):

1. Place 100 mL of the low standard into a 150 mL 

beaker. Add 2 mL of ISA. Stir thoroughly.

2. Rinse electrode with Dl water, blot dry and place in 

the beaker. Wait for a stable reading, and then 

adjust the measuring device to display the value of 

the standard. Refer to the software on-line help or 

the measuring device’s instruction manual for the 

adjustment procedure.

3. Measure 100 mL of the high standard into a second 

150 mL beaker. Add 2 mL of ISA and stir.

4. Rinse electrode with Dl water, blot dry and place in 

the second beaker. Wait for a stable reading, and 

then adjust the measuring device to display the 

value of the second standard.

5. Pipette 100 mL of sample into a 150 mL beaker. Add 

2 mL of ISA. Stir thoroughly.

6. Rinse electrode with Dl water, blot dry and place in 

the sample beaker. Wait for a stable reading and the 

sample concentration will be displayed on the mea

-

suring device.

7. Determine the sample concentration using the 

calibration curve prepared in Step 6 above.

Electrode Storage

Short Term (over night or the weekend):

Rinse the electrode thoroughly with Dl water and place 

the tip in a diluted standard solution between 

measurements. Slide the FastFil sleeve upward to close 

the fill hole.

Refill the reference before putting the electrode back into 

service

Long Term

Refill the electrode with Reference Filling Solution and 

close the fill hole. Remove the sensing element and 

store in the glass vial or replace the cap to protect the 

sensing element.

Follow procedures in "Electrode Preparation" and 

"Checking Electrode Operation" before using the 

electrode again.

Electrode Cleaning

Cleaning should only be attempted if troubleshooting 

methods fail. The PVC membrane is a delicate sensor 

and should not be brushed or otherwise contacted. It 

can be rinsed vigorously under warm water to remove 

debris. Soaking for 10-15 minutes in DI water might be 

useful in extracting other contaminants.

Recondition the electrode by soaking in a low standard 

solution immediately after any cleaning method.

Replacement Items

SC-3523 Ammonium ISE Solutions: 60 mL Standard 

Solution, 30 mL Reference Fill Solution, 30 mL ISA 

Solution

PS-3548 Ammonium ISE Replacement Module

Technical Support

For assistance with any PASCO product, contact 

PASCO at:

Limited Warranty

For a description of the product warranty, see the PASCO catalog.

Copyright

This PASCO scientific instruction sheet is copyrighted with all rights 

reserved. Permission is granted to non-profit educational institutions for 

reproduction of any part of this manual, providing the reproductions are 

used only in their laboratories and classrooms, and are not sold for profit. 

Reproduction under any other circumstances, without the written consent 

of PASCO scientific, is prohibited. Rev: 10/17