Ammonium ISE Probe

Introduction

2

012-15718A

Introduction

This Ammonium Ion Selective Electrode (ISE) Probe is a 

combination electrode that includes both reference and 

sensing half cells in one body housing. It is used for 

measuring ammonium ion (NH

4

+

) activity in aqueous 

samples. 

Technology: Replaceable Polymer/PVC Membrane 

(Ammonium ISE Replacement Module PS-3548)

Overview

Applications

Very popular for environmental, industrial, agriculture, 

food/beverage, and wastewater measurements

Required Equipment

PASCO Wireless pH/ISE/ORP Sensor (PS-3204)

or 

PASPORT High Precision pH/Temperature with 

ISE/ORP Amplifier (PS-2147) with a PASCO 

interface (see www.pasco.com for information)

or

Ion meter or pH/mV meter

PASCO data collection software (see 

www.pasco.com)

Wash bottle filled with distilled or de-ionized water 

Lab wipes 

Pipettes: 1 mL, 10 mL and 100 mL

Clean beakers*

*SE-7287 100 mL Beakers (12 pack) or SE-7288 1000 

mL Beakers (6 pack)

Included Required Solutions*

Reference Fill Solution

1000 ppm Ammonium Standard Solution

Ionic Strength Adjuster (ISE)

(*For replacement items, see the Technical Support sec

-

tion at the end of the document.)

Electrode Preparation

Replace the Module Blank

This electrode performs better when the reference 

junction is kept wetted during transit and storage. 

However, the sensor module is best stored and 

transported dry. Therefore, the electrode is shipped with 

a plastic Module Blank inside that must be removed prior 

to use. 

1. Unscrew the cap of the storage solution bottle (also 

known as “Soaker Bottle”) and remove the 

electrode. Unscrew the plastic Module Blank from 

the end of the electrode. Replace the blank with the 

included Ammonium ISE Replacement Module. 

The Module Blank can be retained for long term 

storage. 

Caution: Do not touch the PVC sensor membrane 

on the end of the Ammonium ISE Replacement 

Module with your fingers.

2. Thoroughly rinse the electrode with distilled water. 

Wipe carefully with a clean lab wipe.

3. Return the electrode to the storage bottle when it is 

not in use.

Fill the Electrode with Electrolyte

The reference (outer) chamber must be filled with 

Reference Fill Solution and remain open during testing:

Size and Material

Body (Epoxy): 12 mm OD x 155 

mm Length

Cap (ABS): 16 mm OD x 57 mm 

Length

Reference

Double-junction Ag/AgCl, ceramic 

pin junction, refillable, sodium 

chloride electrolyte

Features

Replaceable Sensor Module

Range

0.02 to 18,000 mg/L

Reproducibility

±2%

Slope

54-56 mV/decade @ 25°C typical

pH Range

1 to 12

Operating Temperature 0 to 40°C

Interferences

pH<2, Li

+

, Na

+

, K

+

, Cs

+

, Mg

3+

Ca

2+

, Sr

2+

, Ba

2+