DataStudio Calibration

Setup and Calibration

Calibration Procedure:

PS-2108 Dissolved Oxygen Sensor

Equipment required:

• Soaker bottle (included with sensor)
• Calibration Table  (on reverse side)
• Barometric pressure reading for your location
• DataStudio software, version 1.5.3, or a PASPORT datalogger

Equilibrate the Probe in 100% Humidified Air

1. Before performing any calibration, place 5 mL of deionized water into a clean 

soaker bottle.

2. Insert the probe into the soaker bottle and screw on the lid. 

Place the end of the probe about 2 cm above the water (see Figure 1).

3. Shake the soaker bottle vigorously for a few seconds. Shake off any large water 

drops from the membrane.

If using a computer, perform a single-point calibration using the 
DataStudio software. Select a measurement unit, either mg/L or Per-
cent (%) Saturation, before performing the calibration:

Perform a mg/L calibration: 

                                                                                  

1.  Make sure the mg/L checkbox is checked in the PASPORT setup window.
2.  Click the 

Calibrate

 button next to 

mg/L

.

3.  Refer to the calibration table (on the reverse side of this card)  and enter the       

appropriate dissolved oxygen value for the temperature and barometric 
pressure at your location.  You can use a PASPORT Temperature and 
Barometer to take the necessary measurements.  NOTE:  If you call the 
weather service, the barometric pressure they provide must be altitude 
adjusted!

4.  Click the 

Set

 button.

If using a PASPORT datalogger, then:

1.  Turn on the datalogger.
2.  Plug the sensor into the datalogger.
3.  Press the 

Display

 button until ‘Calibrate’ appears on the screen.

4.  Press the 

+

 button to select either mg/L or % as the measurement unit.

5.  Press the 

Check

 button.

For a mg/L calibration

:

•  Refer to the calibration table, enter the appropriate dissolved oxygen value for 

the temperature and barometric pressure at your location (see above for 
detail).  Note, you must have this value to proceed. 

•  Press the 

Tab

  button to move through the digits.

•  Use the 

-

 and 

+

 buttons to set the desired value for each digit.

•  Press the 

Check

 button.

•  Press the 

Display

 button.

For a % saturation calibration

:

•  Wait for the reading to stabilize.  Click the 

Check 

button for 100% saturation.

PASPORT Datalogger Calibration

DataStudio Calibration

Setup and Calibration

Calibration Procedure:

PS-2108 Dissolved Oxygen Sensor

Equipment required:

• Soaker bottle (included with sensor)
• Calibration Table  (on reverse side)
• Barometric pressure reading for your location
• DataStudio software, version 1.5.3, or a PASPORT datalogger

Equilibrate the Probe in 100% Humidified Air

1. Before performing any calibration, place 5 mL of deionized water into a clean 

soaker bottle.

2. Insert the probe into the soaker bottle and screw on the lid. 

Place the end of the probe about 2 cm above the water (see Figure 1).

3. Shake the soaker bottle vigorously for a few seconds. Shake off any large water 

drops from the membrane.

If using a computer, perform a single-point calibration using the 
DataStudio software. Select a measurement unit, either mg/L or Per-
cent (%) Saturation, before performing the calibration:

Perform a mg/L calibration: 

                                                                                  

1.  Make sure the mg/L checkbox is checked in the PASPORT setup window.
2.  Click the 

Calibrate

 button next to 

mg/L

.

3.  Refer to the calibration table (on the reverse side of this card)  and enter the       

appropriate dissolved oxygen value for the temperature and barometric 
pressure at your location.  You can use a PASPORT Temperature and 
Barometer to take the necessary measurements.  NOTE:  If you call the 
weather service, the barometric pressure they provide must be altitude 
adjusted!

4.  Click the 

Set

 button.

If using a PASPORT datalogger, then:

1.  Turn on the datalogger.
2.  Plug the sensor into the datalogger.
3.  Press the 

Display

 button until ‘Calibrate’ appears on the screen.

4.  Press the 

+

 button to select either mg/L or % as the measurement unit.

5.  Press the 

Check

 button.

For a mg/L calibration

:

•  Refer to the calibration table, enter the appropriate dissolved oxygen 

value for the temperature and barometric pressure at your location (see 
above for detail).  Note, you must have this value to proceed. 

•  Press the 

Tab

  button to move through the digits.

•  Use the 

-

 and 

+

 buttons to set the desired value for each digit.

•  Press the 

Check

 button.

•  Press the 

Display

 button.

For a % saturation calibration

:

•  Wait for the reading to stabilize.  Click the 

Check 

button for 100% 

saturation.

PASPORT Datalogger Calibration

012-07688B

012-07688B