background image

Tro u b l e s h o o t i n g

If the Dissolved Oxygen Sensor does not give the 
expected output, perform the three maintenance 
procedures described on this card, checking sensor 
operation after each procedure:

1. “Changing the Electrolyte Solution”
2. “Replacing the Membrane”
3. “Replacing the O-ring”

If the Dissolved Oxygen Sensor still does not function 
properly, contact PASCO Technical Support.

S t o r a g e

Short-Term Storage (two weeks or less)

1. Store the probe in the plastic zip-lock bag with the tip inserted 

in the soaker bottle containing about 10 ml of deionized water.

Long-Term Storage (more than two weeks)

1. Repeat steps 1–3 under “Changing the Electrolyte Solution.”
2. Rinse the soaker bottle with tap water and shake dry.
3. Replace the cartridge housing, and store the probe dry in the 

zip-lock bag with the end of the probe in the soaker bottle to 
protect the membrane from trauma.

R e p l a c i n g   t h e   O - r i n g

The O-ring rarely requires replacing. However, if it develops 
nicks and splits or leaks electrolyte solution from the probe, 
replace it as follows:

1. Follow steps 1–2 under “Replacing the Membrane.”
2. After removing the membrane cartridge, remove the O-ring 

with a pair of fine-tipped tweezers and insert a new O-ring.

3. Reassemble the unit as directed in steps 4–7 of 

Changing

 

the Electrolyte Solution.”

Tro u b l e s h o o t i n g

If the Dissolved Oxygen Sensor does not give the 
expected output, perform the three maintenance 
procedures described on this card, checking sensor 
operation after each procedure:

1. “Changing the Electrolyte Solution”
2. “Replacing the Membrane”
3. “Replacing the O-ring”

If the Dissolved Oxygen Sensor still does not function 
properly, contact PASCO Technical Support.

S t o r a g e

Short-Term Storage (two weeks or less)

1. Store the probe in the plastic zip-lock bag with the tip inserted 

in the soaker bottle containing about 10 ml of deionized water.

Long-Term Storage (more than two weeks)

1. Repeat steps 1–3 under “Changing the Electrolyte Solution.”
2. Rinse the soaker bottle with tap water and shake dry.
3. Replace the cartridge housing, and store the probe dry in the 

zip-lock bag with the end of the probe in the soaker bottle to 
protect the membrane from trauma.

R e p l a c i n g   t h e   O - r i n g

The O-ring rarely requires replacing. However, if it develops 
nicks and splits or leaks electrolyte solution from the probe, 
replace it as follows:

1. Follow steps 1–2 under “Replacing the Membrane.”
2. After removing the membrane cartridge, remove the O-ring 

with a pair of fine-tipped tweezers and insert a new O-ring.

3. Reassemble the unit as directed in steps 4–7 of 

Changing

 

the Electrolyte Solution.”

Отзывы: