Page 8-2
MiniCapt Mobile Microbial Air Sampler Operations Manual
Chapter 8 Cleaning Procedures
Runtime Cleaning Guidelines
Runtime Cleaning Guidelines
• Remove the impactor head from the defined sampling location.
• Ensure no residuals, e.g. from agar, are present in the impactor head.
• Wash the whole stainless steel surface, including mechanical parts, using an appropriate
detergent solution.
• For example, Veltek Vai Dec-Clean, appropriately diluted as recommended by the
detergent manufacturer data sheet. Typically 0.7 Vol% (1 fl oz/Gallon) in WFI water.
• Make sure the impactor head is well cleaned and any cleaning agent residues are
evaporated before proceeding with the autoclave sterilization process.
The above procedures are intended to provide general guidelines and may not be applicable
in some specific manufacturing scenarios.