Chapter 4 Configuring for Ethernet
MiniCapt Remote Microbial Air Sampler Operations Manual
Page 4-3
When communications are established with the unit, a prompt appears. Enter the
Configuration commands (see “Configuration Commands” below) to set the following
addresses:
4
Set the IP address.
5
Set the Net mask.
6
Set the gateway address.
7
Enter the “write” command.
Configuration Commands
?
Displays a command summary for the selected operational mode (a summary of the
information in this table).
Sta(tus)
Displays the following information:
•
Current firmware version number
•
Serial number
•
MAC address
•
IP address
•
Net mask
•
Gateway
•
Queue size (applicable for PMS Ethernet and Modbus TCP connections)
•
Note
: The RS-485 has its own queue fixed to 10.
•
Flow rate
•
Compressed gas mode
•
Connection status (if used with Modbus)
•
Operational mode (Sampling/Idle)
Set ip aaa.bbb.ccc.ddd
The IP address is used when communicating across networks. This command sets the
address in the form of aaa.bbb.ccc.ddd. Each three-digit series is a value of 0–255,
separated by a period (
.
) character. An example of a valid IP address is
010.255.000.060
.
Set mas(k) aaa.bbb.ccc.ddd
Sets the network mask address using a decimal notation form of aaa.bbb.ccc.ddd.
Each three-digit series is a value of 0-255 separated by a period (.) character. The mask
separates the network address from the host address. This address is common to all
devices on the logical network. Usually the mask will be similar to 255.255.255.000.
Your network administrator will tell you if you need to enter this field.