Compressed Gases Kit
Chapter 6 Accessories
MiniCapt Remote Microbial Air Sampler Operations Manual
Page 6-5
>> To install the Compressed Gas Kit on the MiniCapt:
1
Follow the instructions in
on page 6-2 for installing the agar
plate.
2
on page 6-4 as needed for the following steps.
3
Install the Dual Connection BioCapt Impactor Head over the agar plate.
4
Install the Diffusion Chamber on top of the Dual Connection BioCapt Impactor Head.
5
Connect the Solenoid valve tubing to the
Compressed Gas Port
on the bottom of the
MiniCapt Remote Microbial Air Sampler (see
).
6
Connect the Solenoid cable to the
Compressed Gas Solenoid Connector
on the
bottom of the MiniCapt Remote Microbial Air Sampler (see
).
Compressed
Gas Port
Compressed Gas
Solenoid Connector
Figure 6-7 Compressed Gas Solenoid
connector on the bottom of the
MiniCapt Remote Microbial Air Sampler
7
Locate the short piece of tubing that came in the Compressed Gases Kit:
a.
Connect one end of the tubing to the
Solenoid Valve
.
b.
Connect the Solenoid valve to the
Regulator
.